Preliminary Evaluation of a Combined MicroPET®-MR System
Open Access
- 1 February 2010
- journal article
- research article
- Published by SAGE Publications in Technology in Cancer Research & Treatment
- Vol. 9 (1) , 53-60
Abstract
There are many motivations for adding simultaneously acquired MR images to PET scanning. The most straight forward are, superior registration of MR and PET images, the addition of morphological detail when there is non-rigid motion and for pre-clinical studies simultaneous imaging could lead to a significant reduction in the time that animals are required to be anesthetised. In addition simultaneous MR has the potential to provide accurate motion correction for PET image reconstruction. For functional imaging simultaneous acquisition is required to assess the subject in the same physiological state, such as acute stroke studies. The elimination of the additional radiation associated with combining CT with PET, by providing anatomic detail with MR, would be a crucial advantage for cancer screening. Combining the two instruments necessitates some engineering tradeoffs, especially associated with the use of the highly developed photomultiplier tube (PMT) used for light amplification, because of its incompatibility with strong magnetic fields. Our approach is to provide a split in the magnet and gradients to locate the magnetic sensitive components, the PMTs, in regions of low magnetic field, leaving only the essential PET components, the scintillator blocks, in the strong magnetic field region. The crystals are coupled to the PMTs by extending the optical fibres. A further advantage accrues by moving the PET electronics out of the region seen by the MR radio-frequency (RF) and gradient coils as electromagnetic interference effects between the PET and MR systems, which could cause artefacts in either modality, are eliminated. Here we describe a preliminary evaluation of the system, which is essentially a microPET Focus-120 located in a 1T split magnet, and compare its performance to previous microPET instruments.Keywords
This publication has 16 references indexed in Scilit:
- Split gradient coils for simultaneous PET‐MRIMagnetic Resonance in Medicine, 2009
- PET/MRI system designEuropean Journal of Nuclear Medicine and Molecular Imaging, 2008
- Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodesPhysics in Medicine & Biology, 2006
- Performance evaluation of the microPET R4 PET scanner for rodentsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2003
- Studies with a prototype high resolution PET scanner based on LSO-APD modulesIEEE Transactions on Nuclear Science, 1998
- Exact and approximate rebinning algorithms for 3-D PET dataIEEE Transactions on Medical Imaging, 1997
- Initial results from the Sherbrooke avalanche photodiode positron tomographIEEE Transactions on Nuclear Science, 1996
- Automatic, localized in Vivo adjustment of all first‐and second‐order shim coilsMagnetic Resonance in Medicine, 1993
- Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents countsIEEE Transactions on Nuclear Science, 1990
- Analytic 3D image reconstruction using all detected eventsIEEE Transactions on Nuclear Science, 1989