Behavior of Gaseous Microemboli in Extracorporeal Circuits: Air versus Co2
Authors:
- 1 June 2006
- journal article
- research article
- Published by SAGE Publications in The International Journal of Artificial Organs
- Vol. 29 (6) , 578-582
Abstract
Background: Open heart surgery is associated with serious risk of cerebral and peripheral organ dysfunction, attributed in part to air microbubbles generated in or not eliminated from the extracorporeal circuit (ECC). Venous air leakage leads to increased arterial bubble load. CO2 replacing air in cardiac chambers show faster resorption times, reducing possible cerebral or peripheral organ damage after heart valve interventions. In two models of ECC the effect of air entering closed circuits was demonstrated and compared to the effect of CO2 entry. Methods: Fragmentation and dissolution of gas (0.5 mL) was evaluated in an in vitro model of ECC, using ultrasonic bubble detection. Air leakage (10 mL) in the venous line of the ECC was simulated and compared to the effect of the same quantity of CO2 entering the circuit. Both models used whole blood priming and physiological conditions, verified with blood gas analyses. Results: Fragmentation and dissolution of gas bubbles injected into a closed ECC could be demonstrated, complete resorption of CO2 bubbles was observed earlier than complete resorption of room air (5.0±0.6 vs. 14.4±5.9 min, p=0.0009). CO2 entering the venous line lead to 40% less arterial bubble load as compared to the same amount of room air entering the circuit (2099±991 vs. 3555±632, p=0.005). Conclusions: Current ECC systems fail to eliminate gas entering the circuit, leading rather to microbubble dispersion. CO2 is much faster resorbed within the circuit than room air. In open heart surgery as valvular operations, CO2 insufflation into the operative field is protective, as we have demonstrated in our models.Keywords
This publication has 17 references indexed in Scilit:
- Improved cerebral protection through replacement of residual intracavital air by carbon dioxide: a porcine model using diffusion-weighted magnetic resonance imagingThe Journal of Thoracic and Cardiovascular Surgery, 2004
- Improved methods for measurement of gaseous microbubbles during extracorporeal circulationPerfusion, 2002
- Neuropsychologic impairment after coronary bypass surgery: Effect of gaseous microemboli during perfusionist interventionsThe Journal of Thoracic and Cardiovascular Surgery, 2001
- Carbon dioxide digital subtraction angiographyVasa, 1999
- Carbon Dioxide Field Flooding Minimizes Residual Intracardiac Air After Open Heart OperationsThe Annals of Thoracic Surgery, 1997
- Cognitive Outcome After Coronary Artery Bypass: A One-Year Prospective StudyThe Annals of Thoracic Surgery, 1997
- Estimate of the maximum absorption rate of microscopic arterial air emboli after entry into the arterial circulation during cardiac surgeryPerfusion, 1996
- Postoperative management of cerebral air embolism: Gas physiology for surgeonsThe Annals of Thoracic Surgery, 1995
- When do cerebral emboli appear during open heart operations? A transcranial Doppler studyThe Annals of Thoracic Surgery, 1991
- Carbon dioxide in the prevention of air embolism during open-heart surgeryThorax, 1968