Simultaneous quantitation of etoricoxib, salicylic acid, valdecoxib, ketoprofen, nimesulide and celecoxib in plasma by high‐performance liquid chromatography with UV detection
Authors:
- 12 July 2005
- journal article
- research article
- Published by Wiley in Biomedical Chromatography
- Vol. 20 (1) , 125-132
Abstract
A specific, accurate, precise and reproducible high performance liquid chromatography (HPLC) method was developed and validated for the simultaneous quantitation of etoricoxib, salicylic acid, valdecoxib, ketoprofen, nimesulide and celecoxib in human plasma. The method employed a simple liquid_liquid extraction of etoricoxib, salicylic acid, valdecoxib, ketoprofen, nimesulide and celecoxib and internal standard (IS, DRF‐4367) from human plasma (500 µL) into acetonitirile. The organic layer was separated and evaporated under a gentle stream of nitrogen at 40°C. The residue was reconstituted in the mobile phase and injected onto a Kromasil KR 100‐5C18 column (4.6 × 250 mm, 5 µm). The chromatographic separation was achieved by gradient elution consisting of 0.05 M formic acid (pH 3)‒acetonitrile‒methanol‒water at a flow rate of 1.0 mL/min. The eluate was monitored using an ultraviolet (UV) detector set at 235 nm. The ratio of peak area of each analyte to IS was used for quantification of plasma samples. Nominal retention times of etoricoxib, salicylic acid, valdecoxib, ketoprofen, nimesulide, IS and celecoxib were 15.63, 17.20, 21.66, 24.95, 26.27, 30.24 and 32.22 min, respectively. The standard curve for etoricoxib, salicylic acid, valdecoxib, ketoprofen and celecoxib was linear (r2 > 0.999) in the concentration range 0.1‒50 µg/mL and for nimesulide (r2 > 0.999) in the concentration range 0.5‒50 µg/mL. Absolute recovery was >83% from human plasma for all the analytes and IS. The lower limit of quantification (LLOQ) of nimesulide was 0.5 µg/mL and for etoricoxib, salicylic acid, valdecoxib, ketoprofen and celecoxib the LLOQ was 0.1 µg/mL. The inter‐ and intra‐day precisions in the measurement of QC samples, 0.1, 0.3, 15.0 and 40.0 µg/mL (for all analytes except nimesulide), were in the range 2.29‒9.37% relative standard deviation (RSD) and 0.69‒10.28% RSD, respectively. For nimesulide the inter‐ and intra‐day precisions in the measurement of quality control (QC) samples, 0.5, 1.5, 15.0 and 40.0 µg/mL, were in the range 3.21‒7.37% RSD and 0.97‒7.06% RSD, respectively. Accuracy in the measurement of QC samples for all analytes was in the range 91.03‒106.38% of the nominal values. All analytes including IS were stable in the battery of stability studies, viz. bench top, autosampler and freeze_thaw cycles. Stability of all analytes was established for 21 days at −20°C. The application of the assay in an oral pharmacokinetic study in rats co‐administered with celecoxib and valdecoxib is described. Copyright © 2005 John Wiley & Sons, Ltd.Keywords
This publication has 13 references indexed in Scilit:
- A liquid chromatography–mass spectrometry method for the quantification of both etoricoxib and valdecoxib in human plasmaBiomedical Chromatography, 2004
- Simultaneous determination of arylpropionic acidic non-steroidal anti-inflammatory drugs in pharmaceutical formulations and human plasma by HPLC with UV detectionJournal of Pharmaceutical and Biomedical Analysis, 2003
- Screening Procedure for Detection of Non-Steroidal Anti-inflammatory Drugs and their Metabolites in Urine as Part of a Systematic Toxicological Analysis Procedure for Acidic Drugs and Poisons by Gas Chromatography- Mass Spectrometry after Extractive Methylation*Journal of Analytical Toxicology, 2001
- Simultaneous analysis of anthranilic acid derivatives in pharmaceuticals and human urine by high-performance liquid chromatography with isocratic elutionJournal of Chromatography B: Biomedical Sciences and Applications, 2000
- The GC/MS analysis of some commonly used non-steriodal anti-inflammatory drugs (NSAIDs) in pharmaceutical dosage forms and in urineForensic Science International, 1999
- Simultaneous analysis of 14 non-steroidal anti-inflammatory drugs in human serum by electrospray ionization-tandem mass spectrometry without chromatographyJournal of the American Society for Mass Spectrometry, 1998
- Rapid screening for acidic non-steroidal anti-inflammatory drugs in urine by gas chromatography-mass spectrometry in the selected-ion monitoring modeJournal of Chromatography B: Biomedical Sciences and Applications, 1996
- Analytical Methods Validation: Bioavailability, Bioequivalence, and Pharmacokinetic StudiesJournal of Pharmaceutical Sciences, 1992
- Liquid chromatography with amperometric detection for the determination of non-steroidal anti-inflammatory drugs in plasmaJournal of Chromatography B: Biomedical Sciences and Applications, 1990
- Rapid high-performance liquid chromatographic assay for the simultaneous analysis of non-steroidal antiinflammatory drugs in plasmaJournal of Chromatography B: Biomedical Sciences and Applications, 1987