A method to determine diastolic blood pressure based on pressure pulse propagation in the electronic palpation method

Authors:
  • SNS.M. Nissila
    S.M. Nissila
  • PKP.M. Karja-Koskenkari
    P.M. Karja-Koskenkari
  • JKJ.K. Koskenkari
    J.K. Koskenkari
  • MLM.K. Lilja
    M.K. Lilja
  • YKY.A. Kesaniemi
    Y.A. Kesaniemi

Abstract
Evaluates the accuracy of the pulse transit time method for determining diastolic blood pressure using intra-arterial blood pressure as a reference. First, the paper describes the method. Then it discusses two sets of measurements which were carried out to determine the accuracy of the method with cardiac operated patients and healthy young volunteers. The thus obtained accuracies were +0.7 mmHg /spl plusmn/ 10.7 mmHg for the cardiac patients and -6.6 mmHg /spl plusmn/ 10.5 mmHg for the young volunteers, respectively.