Debris of carbon-fibers originated from a CFRP (pEEK) wrist-plate triggered a destruent synovitis in human
- 12 January 2016
- journal article
- case report
- Published by Springer Nature in Journal of Materials Science: Materials in Medicine
- Vol. 27 (3) , 1-10
Abstract
Application of carbon-fiber-reinforced-polymer (CFRP) artifacts in humans has been promoted in Orthopedic and Trauma Surgery. Literature documents the biocompatibility of materials used, namely carbon fibers (CF) and poly-ether thermoplastics, like poly-ether-ether-ketone (PEEK). A properly designed and accurately implanted composite artifact should not expose its fibers during or after surgery: however this may happen. A white Caucasian woman came to our attention 11 months after surgery for a wrist fracture. She had a severe impairment, being unable to flex the thumb; index finger and distal phalanx of third finger. We retrieved a correctly positioned plate and documented an aggressive erosive flexor tendons synovitis with eroded stumps of flexor tendons. The plate and soft tissues were analyzed by Visible Light and Scanning Electron Microscopy. Histopathology showed granulomatous fibrogenic process with CF engulfed inside multinucleated giant cells. Fibers were unmasked and disrupted inside the holes where screws were tightened and corrugation of the polymer coating led to further unmasking. The mechanism of foreign-body reaction to CF has not been studied in depth yet, particularly at the ultrastructural level and in Humans. This case documents a damage occurred in a clinical application and which was theoretically possible. Our opinion is that a proper way to promote the use of CRFP in the Clinic in the short term is to direct Research towards finding a better way to prevent CF debris to be exposed and released. In the longer term, the biological response to CF deserves a deeper understanding.Keywords
This publication has 24 references indexed in Scilit:
- Self-assembling of large ordered DNA arrays using superhydrophobic patterned surfacesNanotechnology, 2013
- Carbon fiber reinforced PEEK Optima—A composite material biomechanical properties and wear/debris characteristics of CF-PEEK composites for orthopedic trauma implantsJournal of the Mechanical Behavior of Biomedical Materials, 2013
- Inflammatory response against different carbon fiber-reinforced PEEK wear particles compared with UHMWPE in vivoActa Biomaterialia, 2010
- Tissue response to the components of a hydroxyapatite‐coated composite femoral implantJournal of Biomedical Materials Research Part A, 2010
- Treatment of painful articular cartilage defects of the patella by carbon-fibre implantsThe Journal of Bone and Joint Surgery. British volume, 1998
- Failure of a Carbon Fiber ImplantSpine, 1998
- In vitro biocompatibility testing of polymers for orthopaedic implants using cultured fibroblasts and osteoblastsBiomaterials, 1995
- Interbody Lumbar Fusion Using a Carbon Fiber Cage Implant Versus Allograft BoneSpine, 1994
- Biological response to chopped‐carbon‐fiber‐reinforced peekJournal of Biomedical Materials Research, 1992
- Carbon fibre composites for orthopaedic implantsComposites, 1978