In-vivo precision of the GE Lunar iDXA densitometer for the measurement of appendicular and trunk lean and fat mass
Authors:
- 30 October 2013
- journal article
- research article
- Published by Springer Nature in European Journal of Clinical Nutrition
- Vol. 67 (12) , 1331-1333
Abstract
Dual energy X-ray absorptiometry (DXA) is considered to be a reference standard for the assessment of body composition, and one particular advantage is its provision of regional body composition parameters. The objective of this study was to determine the in vivo precision of the Lunar iDXA for the measurement of appendicular and trunk lean and fat mass in a heterogeneous group of adults (n=39; age 33.5 (8.3) years; BMI 24.6 (5.4) kg m−2). Two consecutive total body scans were performed on each participant with re-positioning between scans. There was excellent agreement between consecutive scans for measurements of both fat and lean appendicular and trunk composition (R2=0.99). Precision for body composition at all regions was less than 2% coefficient of variation (CV). Precision for bilateral appendicular body composition was less than 4% CV for the arms and less than 2.5% CV for the legs. Our findings indicate that the iDXA is a valuable tool for repeat measurements of regional body composition in adults.Keywords
This publication has 11 references indexed in Scilit:
- Precision of GE Lunar iDXA for the Measurement of Total and Regional Body Composition in Nonobese AdultsJournal of Clinical Densitometry, 2012
- Clinical Observations in Total Body DXA: Technical Aspects of Positioning and AnalysisJournal of Clinical Densitometry, 2012
- In vivo precision of the GE Lunar iDXA densitometer for the measurement of total body composition and fat distribution in adultsEuropean Journal of Clinical Nutrition, 2010
- Dynamics of Change in Total and Regional Body Composition After Gastric Bypass in Obese PatientsObesity, 2010
- Reproducibility of dual-energy x-ray absorptiometry total and regional body composition measurements using different scanning positions and definitions of regionsMetabolism, 2009
- Body Composition Precision with the Lunar iDXAJournal of Clinical Densitometry, 2009
- Body composition by dual‐energy X‐ray absorptiometry in women with previous pre‐eclampsia or small‐for‐gestational‐age offspringBJOG: An International Journal of Obstetrics & Gynaecology, 2009
- Regional fat distribution in adolescents with anorexia nervosa: effect of duration of malnutrition and weight recoveryEuropean Journal of Endocrinology, 2007
- Cross-calibration of GE/Lunar pencil and fan-beam dual energy densitometers—bone mineral density and body composition studiesEuropean Journal of Clinical Nutrition, 2003
- Dual-energy x-ray absorptiometry for total-body and regional bone-mineral and soft-tissue compositionThe American Journal of Clinical Nutrition, 1990