Revisiting the Phadia/EliA cut-off values for anticardiolipin and anti-β2-glycoprotein I antibodies: a systematic evaluation according to the guidelines
- 31 May 2018
- journal article
- research article
- Published by SAGE Publications in Lupus
- Vol. 27 (9) , 1446-1454
Abstract
Background: Phadia/EliA fluorescence enzyme immunoassays are widely used automated assays for anticardiolipin (aCL) and anti-β2-glycoprotein I (aβ2GPI) antibodies. To date, cut-off values for these assays have not been evaluated systematically and the evidence behind manufacturer’s recommended cut-off values is not clear. Objective: To determine Phadia/EliA cut-off values for antiphospholipid antibodies (aPL) according to the procedures suggested by guidelines. Methods: A total of 266 blood donors (135 females and 131 males) were included. The pre-handling and analysis of the samples were performed according to the International Society on Thrombosis and Hemostasis (ISTH) guideline for solid phase aPL assays. Cut-off values and corresponding 90% confidence intervals (CI) for each antibody were established and outliers were handled according to the Clinical and Laboratory Standards Institute (CLSI) guideline for reference intervals. Samples from 377 consecutive patients, referred to our thrombophilia center with evidence of thrombosis or pregnancy morbidity were included for aPL testing. Results: The in-house 99th (97.5th) percentile cut-off values were 11 (8.7), 12 (6.9) 8.5 (5.0) AU/mL for aβ2GPI IgG, IgM and IgA, and 21 (13) GPL-U/mL and 41 (25) MPL-U/mL for aCL IgG and IgM, respectively. The prevalence of positive results (%) defined by these cut-off values in patients with evidence of thrombosis or pregnancy morbidity was 9.5 (12.2), 1.6 (2.9), and 7.0 (9.9), and 0.8 (3.8) for aβ2GPI IgG, IgM, and aCL IgG and IgM respectively. The use of in-house 99th percentile cut-off values compared to the manufacturer suggested cut-off values resulted in 1 and 39 fewer samples for aβ2GPI and aCL to be classified as positive for aPL, respectively. Conclusions: We present Phadia/EliA cut-off values with 90% CI for aPL determined systematically according to the ISTH and CLSI guidelines. These values are different from values previously determined, suggesting variation of aPLs in different populations. Our findings indicate the need for each laboratory to determine/validate assay specific cut-off values for aPL.Keywords
This publication has 25 references indexed in Scilit:
- Anti-cardiolipin and anti-β2-glycoprotein I antibodies: normal reference ranges in northwestern ItalyLupus, 2012
- International consensus guidelines on anticardiolipin and anti–β2‐glycoprotein I testing: Report from the 13th International Congress on Antiphospholipid AntibodiesArthritis & Rheumatism, 2011
- Evaluation of a new set of automated chemiluminescense assays for anticardiolipin and anti-beta2-glycoprotein I antibodies in the laboratory diagnosis of the antiphospholipid syndromeThrombosis Research, 2011
- Update of the guidelines for lupus anticoagulant detectionJournal of Thrombosis and Haemostasis, 2009
- Accuracy of the First Fully Automated Method for Anti‐cardiolipin and Anti‐β2 Glycoprotein I Antibody Detection for the Diagnosis of Antiphospholipid SyndromeAnnals of the New York Academy of Sciences, 2009
- Prevalence of Anticardiolipin Antibodies in Patient Cohorts with Distinct Clinical Manifestations of the Antiphospholipid SyndromeAnnals of the New York Academy of Sciences, 2009
- European attempts for the standardisation of the antiphospholipid antibodiesLupus, 2009
- Influence of different IgG anticardiolipin antibody cut‐off values on antiphospholipid syndrome classificationJournal of Thrombosis and Haemostasis, 2008
- Consensus guidelines on anti-beta 2 glycoprotein I testing and reportingPathology, 2008
- International consensus statement on an update of the classification criteria for definite antiphospholipid syndrome (APS)Journal of Thrombosis and Haemostasis, 2006