Sensitivity of GC‐EI/MS, GC‐EI/MS/MS, LC‐ESI/MS/MS, LC‐Ag+CIS/MS/MS, and GC‐ESI/MS/MS for analysis of anabolic steroids in doping control
Authors:
- 21 October 2015
- journal article
- research article
- Published by Wiley in Drug Testing and Analysis
- Vol. 7 (11-12) , 1040-1049
Abstract
This study compared the sensitivity of various separation and ionization methods, including gas chromatography with an electron ionization source (GC‐EI), liquid chromatography with an electrospray ionization source (LC‐ESI), and liquid chromatography with a silver ion coordination ion spray source (LC‐Ag+CIS), coupled to a mass spectrometer (MS) for steroid analysis. Chromatographic conditions, mass spectrometric transitions, and ion source parameters were optimized. The majority of steroids in GC‐EI/MS/MS and LC‐Ag+CIS/MS/MS analysis showed higher sensitivities than those obtained with other analytical methods. The limits of detection (LODs) of 65 steroids by GC‐EI/MS/MS, 68 steroids by LC‐Ag+CIS/MS/MS, 56 steroids by GC‐EI/MS, 54 steroids by LC‐ESI/MS/MS, and 27 steroids by GC‐ESI/MS/MS were below cut‐off value of 2.0 ng/mL. LODs of steroids that formed protonated ions in LC‐ESI/MS/MS analysis were all lower than the cut‐off value. Several steroids such as unconjugated C3‐hydroxyl with C17‐hydroxyl structure showed higher sensitivities in GC‐EI/MS/MS analysis relative to those obtained using the LC‐based methods. The steroids containing 4, 9, 11‐triene structures showed relatively poor sensitivities in GC‐EI/MS and GC‐ESI/MS/MS analysis. The results of this study provide information that may be useful for selecting suitable analytical methods for confirmatory analysis of steroids. Copyright © 2015 John Wiley & Sons, Ltd.Keywords
This publication has 21 references indexed in Scilit:
- Analytical methods for anti-doping control in sport: anabolic steroids with 4,9,11-triene structure in urineTrAC Trends in Analytical Chemistry, 2011
- High sensitive analysis of steroids in doping control using gas chromatography/time‐of‐flight mass‐spectrometryDrug Testing and Analysis, 2010
- Feasibility of gas chromatography–microchip atmospheric pressure photoionization-mass spectrometry in analysis of anabolic steroidsJournal of Chromatography A, 2010
- Ionization of anabolic steroids by adduct formation in liquid chromatography electrospray mass spectrometryJournal of Mass Spectrometry, 2007
- Mass spectrometry in sports drug testing: Structure characterization and analytical assaysMass Spectrometry Reviews, 2006
- Fast screening of anabolic steroids and other banned doping substances in human urine by gas chromatography/tandem mass spectrometryJournal of Mass Spectrometry, 2002
- Liquid chromatography/mass spectrometry in anabolic steroid analysis—optimization and comparison of three ionization techniques: electrospray ionization, atmospheric pressure chemical ionization and atmospheric pressure photoionizationJournal of Mass Spectrometry, 2002
- Influence of a ring substituent on the tendency to form H2O adducts to Ag+ complexes with phenylalanine analogues in an ion trap mass spectrometerJournal of Mass Spectrometry, 2002
- Screening and confirmation analysis of anabolic agents in human urine by gas chromatography — hybrid mass spectrometry (high resolution — time of flight)Analytica Chimica Acta, 2001
- Mechanistic aspects of fatty acid retention in silver ion chromatographyJournal of Chromatography A, 1996