Technical assessment of a novel vertical CT system for upright radiotherapy simulation and treatment planning
Preprint
Abstract
Purpose: To characterize image quality, imaging dose, and dose calculation accuracy for an upright CT scanner with a six-degree-of-freedom patient positioning system. Methods: Imaging dose (CTDIvol) was measured at 120 kVp and 200 mAs. Image quality was evaluated using an ACR-464 phantom. Mean CT number accuracy was assessed within inserts of known material and uniformity as the difference in values at the center and periphery of uniform phantoms. High-contrast resolution was assessed by visible line pairs and modulation transfer function (MTF). Low-contrast performance was quantified by contrast-to-noise-ratio (CNR). Spatial integrity was evaluated between fiducials 100 mm apart. Hounsfield unit to mass density and stopping-power-ratio calibrations were performed. Proton and photon treatment plans were optimized on upright CT scans of a thorax phantom in heterogenous and homogeneous regions. Dose was forward computed on a registered recumbent CT scan and agreement evaluated using 3D gamma analysis. Results: CT imaging dose (CTDIvol) was 23.5 mGy for the 16 cm head phantom and 10.1 mGy for the 32 cm body phantom. Mean CT numbers (HU) were within the expected range for water (1.7) and acrylic (120.8). CT numbers were slightly (5-27 HU) out-of-range for air (-950.4), polyethylene (-78.8), and bone (823.0). Image uniformity was 20.2 HU and 35.0 HU for 20 cm and 48 cm diameter phantoms, respectively. Eight high-contrast line pairs were visualized. The MTF equaled 4.4 cm-1 at 50% and 7.1 cm-1 at 10%. The median CNR was 0.93, below the 1.0 tolerance. Spatial integrity was 0.36 mm. Gamma pass rates were 99.8% for photon and 90.6% for proton plans with 1%/1mm criteria, and greater than or equal to 98.0% for all plans with 3%/2mm criteria. Conclusion: Upright CT image quality and dose calculation accuracy are acceptable for photon and proton radiotherapy.Keywords
All Related Versions
This publication has 40 references indexed in Scilit:
- A radiation therapy platform to enable upright cone beam computed tomography and future upright treatment on existing photon therapy machinesMedical Physics, 2024
- Calculation method for novel upright CT scanner isodose curvesJournal of Applied Clinical Medical Physics, 2024
- Relative thoracic changes from supine to upright patient position: A proton collaborative group studyJournal of Applied Clinical Medical Physics, 2023
- Anatomical changes in the male pelvis between the supine and upright positions—A feasibility study for prostate treatments in the upright positionJournal of Applied Clinical Medical Physics, 2023
- Assessing Equity of Access to Proton Beam Therapy: A Literature ReviewClinical Oncology, 2023
- Estimating right atrial pressure using upright computed tomography in patients with heart failureEuropean Radiology, 2022
- Development of an isocentric rotating chair positioner to treat patients of head and neck cancer at upright seated position with multiple nonplanar fields in a fixed carbon‐ion beamlineMedical Physics, 2020
- Tolerance limits and methodologies forIMRTmeasurement‐based verificationQA:Recommendations ofAAPMTask Group No. 218Medical Physics, 2018
- Reproducibility of patient setup in the seated treatment position: A novel treatment chair designJournal of Applied Clinical Medical Physics, 2017
- An isocentric chair for the simulation and treatment of radiation therapy patientsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1991