Retention and removal of a new viscous dispersive ophthalmic viscosurgical device during cataract surgery in animal eyes
Authors:
- T OshikaTOT Oshika
- F OkamotoFOF Okamoto
- Y KajiYKY Kaji
- T HiraokaTHT Hiraoka
- T KiuchiTKT Kiuchi
- M SatoMSM Sato
- K KawanaKKK Kawana
- 1 April 2006
- journal article
- research article
- Published by BMJ in British Journal of Ophthalmology
- Vol. 90 (4) , 485-487
Abstract
Aims: To assess the retention and removal properties of a new viscous dispersive ophthalmic viscosurgical device (OVD), DisCoVisc, in comparison with those of cohesive (Provisc), dispersive (Viscoat), and viscoadaptive (Healon5) OVDs. Methods: In 20 porcine eyes, cataract surgery was simulated using one of the four OVDs which were stained with fluorescein for better visualisation. Three parameters were measured. Firstly, the presence/absence of OVDs in the chamber at the completion of phacoemulsification was recorded. Secondly, the time until the OVDs were completely removed from the anterior chamber using the phaco needle was measured. Thirdly, after intraocular lens (IOL) implantation, the time needed to completely remove the OVDs from the chamber with irrigation/aspiration tip was recorded. Results: At the completion of phacoemulsification, the OVDs retained in 0% (0/5) for Provisc, 80% (4/5) for Healon5, 100% (5/5) for DisCoVisc, and 100% (5/5) for Viscoat. The retention of OVDs during phacoemulsification was greatest with Viscoat followed by, in descending order, DisCoVisc, Healon5, and Provisc. The removal of OVDs after IOL implantation took longest with Viscoat followed by Healon5, DisCoVisc, and Provisc. Conclusion: The viscous dispersive DisCoVisc showed excellent retention during phacoemulsification, while its removal after IOL implantation was very easy. When compared with the viscoadaptive Healon5, DisCoVisc was retained better in the chamber and was easier to remove. These features of DisCoVisc should be highly advantageous when considering covering the entire cataract surgery procedure with a single OVD.Keywords
This publication has 13 references indexed in Scilit:
- Removal times for a dispersive and a cohesive ophthalmic viscosurgical device correlated with intraocular lens materialJournal of Cataract & Refractive Surgery, 2004
- Removal times and techniques of a viscoadaptive ophthalmic viscosurgical deviceJournal of Cataract & Refractive Surgery, 2004
- Healon5: Comparison of 2 removal techniquesJournal of Cataract & Refractive Surgery, 2002
- Clinical results of phacoemulsification with the use of Healon5 or ViscoatJournal of Cataract & Refractive Surgery, 2001
- Healon5 viscoadaptive formulation: Comparison to Healon and Healon GVJournal of Cataract & Refractive Surgery, 2001
- Two-compartment technique to remove ophthalmic viscosurgical devicesJournal of Cataract & Refractive Surgery, 2000
- Dispersive-cohesive viscoelastic soft shell techniqueJournal of Cataract & Refractive Surgery, 1999
- Viscoelastic adherence to corneal endothelium following phacoemulsificationJournal of Cataract & Refractive Surgery, 1998
- New method to measure the retention of viscoelastic agents on a rabbit corneal endothelial cell line after irrigation and aspirationJournal of Cataract & Refractive Surgery, 1998
- Removal of viscoelastic materials after experimental cataract surgery in vitroJournal of Cataract & Refractive Surgery, 1992