Abstract
The goal of this work is to develop a technique to measure the x‐ray diffraction signals of breast biopsy specimens. A biomedical x‐ray diffraction technology capable of measuring such signals may prove to be of diagnostic use to the medical field. Energy dispersive x‐ray diffraction measurements coupled with a semianalytical model were used to extract the differential linear scattering coefficients of breast tissues on absolute scales. The coefficients describe the probabilities of scatter events occuring per unit length of tissue per unit solid angle of detection. They are a function of the momentum transfer argument, , where angle and wavelength. The technique was validated by using a diameter polychromatic x‐ray beam incident on a diameter thick sample of water. Water was used because good x‐ray diffraction data are available in the literature. The scatter profiles from 6° to 15° in increments of 1° were measured with a thick cadmium zinc telluride detector. A diameter Pb aperture was placed on top of the detector. The target to detector distance was and the duration of each measurement was . Ensemble averages of the results compare well with the gold standard data of A. H. Narten [“X‐ray diffraction data on liquid water in the temperature range 4°C–200°C,” ORNL Report No. 4578 (1970)]. An average 7.68% difference for which most of the discrepancies can be attributed to the background noise at low angles was obtained. The preliminary measurements of breast tissue are also encouraging.

This publication has 41 references indexed in Scilit: