Agreement and repeatability of central corneal thickness measurements made by ultrasound pachymetry and anterior segment optical coherence tomography
Authors:
- Chao-Wen LinCLChao-Wen Lin
- Tsing-Hong WangTWTsing-Hong Wang
- Yu-Hsuan HuangYHYu-Hsuan Huang
- Jehn-Yu HuangJHJehn-Yu Huang
- 1 September 2013
- journal article
- Published by Wolters Kluwer Health in Taiwan Journal of Ophthalmology
- Vol. 3 (3) , 98-102
Abstract
PurposeTo evaluate repeatability within and between examiners and to assess agreement in corneal thickness measurements obtained by anterior segment optical coherence tomography (ASOCT) and ultrasonic pachymetry (USP).MethodsThis was a prospective, observational study. Fifty-one patients who visited a glaucoma service with suspected glaucoma, glaucoma, or cataract were enrolled. Patients with a history of corneal surgery or abnormalities and those who were uncooperative for examinations were excluded. Each patient underwent ASOCT followed by USP, with each test performed by two different examiners. Measurement repeatability was evaluated using intraclass correlation coefficient (ICC) values. Agreement in corneal thickness measurements was evaluated by the Bland and Altman plot method. Linear regression analysis was used to assess the relationship between ASOCT and USP measurements.ResultsUSP and ASOCT corneal thickness measurements revealed high intraexaminer and interexaminer repeatability (ICCs ≥ 0.978). ASOCT corneal thickness measurements of a central 2-mm zone showed higher intraexaminer (ICC = 0.999) and interexaminer (ICC = 0.999) repeatability than USP measurements or ASOCT measurements of the vertex. The 95% limit of agreement (LoA) between the vertex-centered ASOCT and the USP scan was between 3.68 and 24.76 μm. The 95% LoA between the central 2-mm zone ASOCT and the USP scan was between 3.75 and 23.39 μm. Average ASOCT corneal thickness was less than USP measurements, but ASOCT thickness could be converted to USP thickness through linear regression equations.ConclusionBoth devices have good intraexaminer and interexaminer repeatability, though ASOCT has slightly better interexaminer repeatability. ASOCT accurately and reliably measures corneal thickness in a noninvasive mannerKeywords
This publication has 16 references indexed in Scilit:
- Measurement of central corneal thickness by high‐resolution Scheimpflug imaging, Fourier‐domain optical coherence tomography and ultrasound pachymetryActa Ophthalmologica, 2012
- Collagen cross-linkage: a comprehensive review and directions for future researchBritish Journal of Ophthalmology, 2009
- Central and peripheral corneal thickness measured with optical coherence tomography, Scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post–laser in situ keratomileusis eyesJournal of Cataract & Refractive Surgery, 2009
- Comparative Study of Central Corneal Thickness Measurement with Slit-Lamp Optical Coherence Tomography and Visante Optical Coherence TomographyOphthalmology, 2008
- Comparison of Central Corneal Thickness using Anterior Segment Optical Coherence Tomography vs Ultrasound PachymetryAmerican Journal of Ophthalmology, 2008
- Risk Assessment for Ectasia after Corneal Refractive SurgeryOphthalmology, 2008
- Effect of corneal thickness on the agreement between ultrasound and Orbscan II pachymetryJournal of Cataract & Refractive Surgery, 2007
- Corneal Pachymetry Mapping with High-speed Optical Coherence TomographyOphthalmology, 2006
- Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetryEye, 2002
- Increased Corneal Thickness Simulating Elevated Intraocular PressureArchives of Ophthalmology (1950), 1978