Development and In Vivo Testing of a Chronic Percutaneous Prosthesis

Authors:

Abstract
Significant progress has been made in the development of a percutaneous energy transmission system (PETS) designed to provide energy to intracorporeally implanted blood pumps for up to 2 years. The mean survival of our six most recent PETS implantations was 351 ± 33 days. Four of these were explanted electively after a full year of implantation. Sinus formation around each prosthesis was limited by a biomaterial seal at the sinus termination point. The results support the hypothesis that collagen inhibits epidermal downgrowth. Continued development of systems based on this hypothesis is desirable.

This publication has 2 references indexed in Scilit: