Repeatability of Wire and Surface Electrodes in Gait
Authors:
- Kay CernyKCKay Cerny
- Olfat MohammedOMOlfat Mohammed
- 1 May 2003
- journal article
- research article
- Published by Wolters Kluwer Health in American Journal of Physical Medicine & Rehabilitation
- Vol. 82 (5) , 338-344
Abstract
Muscle forces are not directly measurable without invasive methods (i.e., tendon force transducers). Techniques such as dynamic electromyography are therefore required to obtain insight into the role of muscles during motion. There is controversy about the choice of recording electrode type. Surface electrodes are noninvasive and allow recording over a large area yet may allow intramuscular crosstalk. Indwelling electrodes also have been used in gait analysis. This electrode type is able to analyze the electrical activity of small or deep muscles. Despite the advantages of intramuscular electrodes for some applications, this electrode type is often rejected because of the need for skin penetration and for specialized personnel to acquire the data. The reliability of the wire electromyographic signal has also been questioned. The objective of this study was to determine if there were differences in the test-retest reliability of surface vs. intramuscular electrodes in gait analysis. Surface and intramuscular electromyographic data were obtained from the soleus muscle in 18 adults with no history of neuromuscular disease as they performed self-selected speed walking. A statistical criterion (variance ratio) was used to measure the reproduction of phasic patterns of muscle activity with both wire and surface electrodes on repeat-day testing of the soleus muscle. Each electrode type was remarkably consistent. Mean variance ratio values for wire electrodes (mean = 187) were slightly lower than mean surface electromyographic variance ratio values (mean = 199). These differences did not approach statistical significance (P = 0.768). The study results show that the dynamic electromyographic signal obtained with wire electrodes is, at minimum, as repeatable as surface electromyograms.Keywords
This publication has 22 references indexed in Scilit:
- Surface vs. fine-wire electrode ensemble-averaged signals during gaitJournal of Electromyography and Kinesiology, 1995
- Surface and wire EMG crosstalk in neighbouring musclesJournal of Electromyography and Kinesiology, 1994
- A comparison of the triceps surae and residual muscle moments at the ankle during cyclingJournal of Biomechanics, 1991
- Spatial distribution of muscle fibers within the territory of a motor unitMuscle & Nerve, 1990
- Below-knee amputee gait with dynamic elastic response prosthetic feet: A pilot studyJournal of Rehabilitation Research and Development, 1990
- Comparative electromyography of the lower extremity in jogging, running, and sprintingThe American Journal of Sports Medicine, 1986
- Predictive Value of Manual Muscle Testing and Gait Analysis in Normal Ankles by Dynamic ElectromyographyFoot & Ankle, 1986
- Repeatability of phasic muscle activity: Performance of surface and intramuscular wire electrodes in gait analysisJournal of Orthopaedic Research, 1985
- Pathodynamics of Gait Alterations in Cerebral Palsy and the Significance of Kinetic Electromyography in Evaluating Foot and Ankle ProblemsFoot & Ankle, 1983
- Comment on criteria for patient evaluation toolsJournal of Biomechanics, 1981