An Electron Current and Energy Regulator for Mass Spectrometer Ion Sources
Authors:
- 1 November 1954
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 25 (11) , 1083-1088
Abstract
An improved electron current and energy regulator for mass spectrometer ion sources is described. The improvements over previous regulator circuits consist of (a) a two kilocycle filament excitation supply which allows improved filament emission control and (b) a method of controlling the filament to shield voltage over a wide range of electron energies. Data on isotope ratio measurements are provided to show the increased stability over that of previous regulators.Keywords
This publication has 7 references indexed in Scilit:
- An Electron Current Regulator for Mass Spectrometer Ion SourcesReview of Scientific Instruments, 1951
- Improvements in Mass Spectrometers for the Measurement of Small Differences in Isotope Abundance RatiosReview of Scientific Instruments, 1950
- A Simplified Emission Regulator for Mass-Spectrometer Ion SourcesReview of Scientific Instruments, 1949
- Temperature Coefficients for the Mass Spectra of a Number of HydrocarbonsThe Journal of Chemical Physics, 1949
- Effect of Temperature on Mass SpectraThe Journal of Chemical Physics, 1949
- Effect of Temperature on the Mass Spectra of HydrocarbonsThe Journal of Chemical Physics, 1947
- Thermionic Control of an Ionization GaugeReview of Scientific Instruments, 1937