Implant Choice May Reduce the Risk of Early Mechanical Failure in Total Ankle Replacement
- 5 March 2024
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 106 (9) , 767-775
Abstract
Background: Total ankle replacements (TARs) have rapidly advanced in terms of volume, technique, design, and indications. However, TARs are still at risk for early mechanical failure and revision. Prior studies have investigated potential risk factors for failure, but have been limited to smaller series or older implants. This study sought to identify risk factors for early mechanical failure in modern TAR. Methods: This is a retrospective study of a single-institution registry. Five surgeons contributed cases involving patients who underwent a primary TAR with any implant. Implants were grouped on the basis of the type of fixation. The primary outcome was early mechanical failure (revision with component removal for a non-infectious etiology, that is, subsidence, aseptic loosening, and/or malalignment). Logistic regression determined the effects of age, weight, hindfoot arthrodesis, implant type, and radiographic deformity on failure. Results: The 731 included patients had a mean follow-up of 2.7 years. Ten percent (71 patients) had hindfoot arthrodesis. There were 33 mechanical failures (4.5%) at a mean of 1.7 years after the index surgical procedure. Our model demonstrated that hindfoot arthrodesis was associated with 2.7 times greater odds of failure (p = 0.045), every 10 kg of body weight increased the odds of tibial-sided failure by 1.29 times (p = 0.039), and implants with more extensive tibial fixation (stems or keels) lowered the odds of tibial failure by 95% (p = 0.031). Conclusions: In patients with uncontrollable risk factors (hindfoot arthrodesis) or risk factors that may or may not be modifiable by the patient (weight), implants with more robust tibial fixation may be able to reduce the risk of early mechanical failure. Further research is warranted to support efforts to decrease early failure in TAR. Level of Evidence: Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.Keywords
This publication has 40 references indexed in Scilit:
- Are Body Mass Index Cutoffs Creating Racial, Ethnic, and Gender Disparities in Eligibility for Primary Total Hip and Knee Arthroplasty?The Journal of Arthroplasty, 2022
- Reduced Aseptic Loosening With Fully Cemented Short-Stemmed Tibial Components in Primary Cemented Total Knee ArthroplastyThe Journal of Arthroplasty, 2020
- Outcomes Following Total Ankle ArthroplastyOrthopedic Clinics of North America, 2019
- Race, Ethnicity, Sex, and Obesity: Is It Time to Personalize the Scale?Mayo Clinic Proceedings, 2019
- Total ankle replacement design and positioning affect implant-bone micromotion and bone strainsMedical Engineering & Physics, 2017
- Failure modes for total ankle arthroplasty: a statistical analysis of the Norwegian Arthroplasty RegisterArchives of orthopaedic and trauma surgery, 2014
- Outcome of unilateral ankle arthrodesis and total ankle replacement in terms of bilateral gait mechanicsJournal of Orthopaedic Research, 2013
- Total ankle arthroplasty versus ankle arthrodesis. Comparison of sports, recreational activities and functional outcomeInternational Orthopaedics, 2011
- The Salto Total Ankle Arthroplasty: Survivorship and Analysis of Failures at 7 to 11 yearsClinical Orthopaedics and Related Research, 2011
- Patient characteristics differently affect early cup and stem loosening in THA: a case-control study on 7,535 patientsInternational Orthopaedics, 2007