A diamond detector in the dosimetry of high-energy electron and photon beams
Authors:
- Theodor W KaulichTKTheodor W Kaulich
- Fridtjof NüsslinFNFridtjof Nüsslin
- 9 August 1999
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 44 (9) , 2183-2192
Abstract
A diamond detector type 60003 (PTW Freiburg) was examined for the purpose of dosimetry with 4-20 MeV electron beams and 4-25 MV photon beams. Results were compared with those obtained by using a Markus chamber for electron beams and an ionization chamber for photon beams. Dose distributions were measured in a water phantom with the detector connected to a Unidos electrometer (PTW Freiburg). After a pre-irradiation of about 5 Gy the diamond detector shows a stability in response which is better than that of an ionization chamber. The current of the diamond detector was measured under variation of photon beam dose rate between 0.1 and 7 Gy min(-1). Different FSDs were chosen. Furthermore the pulse repetition frequency and the depth of the detector were changed. The electron beam dose rate was varied between 0.23 and 4.6 Gy min(-1) by changing the pulse-repetition frequency. The response shows no energy dependence within the covered photon-beam energy range. Between 4 MeV and 18 MeV electron beam energy it shows only a small energy dependence of about 2%, as expected from theory. For smaller electron energies the response increases significantly and an influence of the contact material used for the diamond detector can be surmised. A slight sublinearity of the current and dose rate was found. Detector current and dose rate are related by the expression i alpha Ddelta, where i is the detector current, D is the dose rate and delta is a correction factor of approximately 0.963. Depth-dose curves of photon beams, measured with the diamond detector, show a slight overestimation compared with measurements with the ionization chamber. This overestimation is compensated for by the above correction term. The superior spatial resolution of the diamond detector leads to minor deviations between depth-dose curves of electron beams measured with a Markus chamber and a diamond detector.Keywords
This publication has 11 references indexed in Scilit:
- Energy and dose rate dependence of a diamond detector in the dosimetry of 4–25 MV photon beamsMedical Physics, 1997
- Dose rate dependence of a PTW diamond detector in the dosimetry of a 6 MV photon beamPhysics in Medicine & Biology, 1994
- Evaluation of a PTW diamond detector for electron beam measurementsPhysics in Medicine & Biology, 1993
- Diamond radiation detectorsDiamond and Related Materials, 1993
- Diamond Detectors in Relative Dosimetry of Photon, Electron and Proton Radiation FieldsRadiation Protection Dosimetry, 1990
- Dosimetry with a diamond operating as a resistorPhysics in Medicine & Biology, 1981
- Evaluation of diamond radiation dosemetersPhysics in Medicine & Biology, 1980
- Diamond dosimeter for x-ray and δ-radiationIEEE Transactions on Nuclear Science, 1977
- Preparation and Characteristics of Natural Diamond Nuclear Radiation DetectorsIEEE Transactions on Nuclear Science, 1975
- Some physical properties of diamondsAdvances in Physics, 1956