The clinical performance of a chemiluminescent immunoassay in detecting anti-cardiolipin and anti-β2 glycoprotein I antibodies. A comparison with a homemade ELISA method

Authors:

Abstract
Fully automated chemiluminescence immunoassays (CLIAs) are emerging technologies for the detection of antiSera were collected from 104 patients with primary APS, 88 seronegative subjects who met the clinical but not the laboratory criteria for APS, and 150 control subjects. IgG/IgM aCL and IgG/IgM anti-β2GPI antibodies were determined in the sera using a CLIA (HemosIL AcuStarCLIA had a significantly lower comparative sensitivity for IgM aCL and IgG/IgM IgG anti-β2GPI antibodies; its comparative specificity was higher with respect to ELISA for IgM aCL and IgM anti-β2GPI antibodies. The two techniques showed a high, significant agreement (p<0.001) and a significant titer correlation (p<0.001). CLIA also detected IgG/IgM aCL and IgG anti-β2GPI antibodies in the seronegative patients. There was a significantly higher prevalence of IgG aCL and IgG anti-β2GPI antibodies (p<0.001 and p=0.01, respectively) in those patients with respect to that in the control population.Despite a lower comparative sensitivity, CLIA showed a higher comparative specificity for some aPL and a good level of agreement and correlation with a homemade ELISA. CLIA also detected some aCL and anti-β2GPI antibodies in the seronegative patients not usually identified by homemade ELISA.

This publication has 17 references indexed in Scilit: