Fracture of tissue dilator in the femoral vein during insertion of temporary dialysis catheter: a stitch in time can save nine
- 20 July 2021
- journal article
- research article
- Published by BMJ in BMJ Case Reports
- Vol. 14 (7) , e243835
Abstract
A 48-year-old man presented to the surgery casualty with 1-day history of broken foreign body during the insertion of dialysis catheter and a failed surgical retrieval. A Doppler ultrasonography of the right groin and lower limb and a noncontrast CT of abdomen and pelvis were performed. Eventhough no intravascular foreign body could be identified on imaging, a decision to re-explore the wound was taken in view of definitive clinical history. A 9.5 cm-long, broken piece of tissue dilator was found inside the right external iliac vein, which was removed through venotomy of the femoral vein. Postoperative recovery was uneventful.Keywords
This publication has 11 references indexed in Scilit:
- Retrieval of Intravascular Fractured Fragment of Tunnelled Double Lumen Catheter in Hemodialysis PatientOpen Access Macedonian Journal of Medical Sciences, 2019
- Retained Guide Wire Migrates into the BrainJournal of Vascular and Interventional Radiology, 2018
- Sonographic detection of a radiolucent object in dialysis vascular accessJournal of Clinical Ultrasound, 2014
- Hemodialysis Access Usage Patterns in the Incident Dialysis Year and Associated Catheter-Related ComplicationsAmerican Journal of Kidney Diseases, 2013
- Percutaneous retrieval of intravascular venous foreign bodies in childrenPediatric Radiology, 2011
- Fracture and Embolization of a Temporary Hemodialysis Femoral CatheterThe Journal of Vascular Access, 2010
- DangerNursing, 2007
- US Medicare will stop paying for preventable errorsCanadian Medical Association Journal, 2007
- Transluminal Removal of an Embolized Venous Catheter Fragment with Simple Means: A Case ReportTherapeutic Apheresis and Dialysis, 2003
- Preventing Complications of Central Venous CatheterizationNew England Journal of Medicine, 2003