Position of different nebulizer types for aerosol delivery in an adult model of mechanical ventilation
Authors:
Open Access
- 10 October 2022
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Medicine
- Vol. 9, 950569
Abstract
Background: The optimal positions of different types of nebulizer for aerosol delivery remain unclear.Methods: Three ICU ventilators employing three types of nebulizer were separately connected to a simulated lung to simulate nebulization during invasive ventilation. Assist/control-pressure control (A/C-PC) mode was utilized, with inspiratory pressure (Pi) set to 12 cmH2O and positive end expiratory pressure (PEEP) set to 5 cmH2O, and with a target Vt of 500 ml. The bias flow of all the ventilators was set to 2 L/min. The three nebulizers were the continuous jet nebulizer (c-JN), the inspiratory synchronized jet nebulizer (i-JN), and the vibrating mesh nebulizer (VMN). The five nebulizer positions were as follows: at the Y-piece (position 1) and 15 cm from the Y-piece (position 2) between the endotracheal tube and the Y-piece, at the Y-piece (position 3) and 15 cm from the Y-piece (position 4) in the inspiratory limb; and at the humidifier inlet (position 5). Aerosols were collected with a disposable filter placed at the simulated lung outlet (n = 3) and were measured by UV spectrophotometry (276 nm). The measurements were compared under different experimental conditions.Results: The aerosol delivery of c-JN, i-JN, and VMN was 5.33 ± 0.49~11.12 ± 0.36%, 7.73 ± 0.76~13.75 ± 0.46% and 11.13 ± 56–30.2 ± 1.63%, respectively. The higher aerosol delivery: for c-JN~Positions 2 (10.95 ± 0.15%), fori-JN~Positions 1 or 2 (12.91 ± 0.88% or 13.45 ± 0.42%), for VMN~Positions 4(29.03 ± 1.08%); the lower aerosol delivery: for c-JN~Positions 1, 3 or 5, fori-JN~Positions 4 or 5, for VMN~Positions 5.The highest aerosol delivery:For c-JN at Position 2 (10.95 ± .15%), for i-JN at Position 1 or 2 (12.91 ± .88% or 13.45 ± .42%), for VMN at Positions 4 (29. 03 ± 1.08%); the lower aerosol delivery: for c-JN at Positions 1, 3 or 5, for i-JN at Positions 4 or 5, for VMN at Positions 5. The highest aerosol deliveryof c-JN was lower than that of i-JN while the VMN was the highest (all P < .05). However, no differences were observed between the highest aerosol delivery with c-JN and the lowest aerosol delivery with i-JN. Similar results were found between the lowest aerosol delivery with VMN and the highest aerosol delivery with c-JN /i-JN in the Avea ventilator. There were no differences in the highest aerosol delivery of each nebulizer among the different ventilators (all p > 0.05).Conclusion: During adult mechanical ventilation, the type and position of nebulizer influences aerosol delivery efficiency, with no differences between ventilators.Keywords
This publication has 27 references indexed in Scilit:
- Aerosol therapy in intensive and intermediate care units: prospective observation of 2808 critically ill patientsIntensive Care Medicine, 2015
- Influence of Inspiratory Flow Pattern and Nebulizer Position on Aerosol Delivery with a Vibrating-Mesh Nebulizer During Invasive Mechanical Ventilation: An in Vitro AnalysisJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2014
- Ventilator-Integrated Jet Nebulization Systems: Tidal Volume Control and Efficiency of SynchronizationRespiratory Care, 2014
- In Vitro Evaluation of Aerosol Delivery by Different Nebulization Modes in Pediatric and Adult Mechanical VentilatorsRespiratory Care, 2014
- Aerosol delivery to intubated patientsExpert Opinion on Drug Delivery, 2013
- Inhalation Therapy in Patients Receiving Mechanical Ventilation: An UpdateJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2012
- Bilevel vs ICU Ventilators Providing Noninvasive Ventilation: Effect of System LeaksChest, 2009
- Aerosol Delivery During Mechanical Ventilation: From Basic Techniques to New DevicesJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2008
- Aerosol Delivery and Modern Mechanical VentilationAmerican Journal of Respiratory and Critical Care Medicine, 2003
- Delivery of a Nebulized Aerosol to a Lung Model during Mechanical Ventilation: Effect of Ventilator Settings and Nebulizer Type, Position, and Volume of FillAmerican Review of Respiratory Disease, 1992