The use of radiochromic film in treatment verification of dynamic stereotactic radiosurgery

Authors:
  • ALA. W. Lightstone
    A. W. Lightstone
  • DMD. L. D. Mason
    D. L. D. Mason
  • POP. F. O'Brien
    P. F. O'Brien

Abstract
Standard silver‐based films are usually too sensitive to be used as direct indicators of dose in dynamic radiosurgery because of optical saturation. This paper describes the use of a new radiochromic film to measure 6‐MV radiosurgery doses and dose distributions in a head phantom. Dose calibration of the radiochromic film was performed in the range of 2.3–50.2 Gy using light of 632‐ and 530‐nm wavelengths. Radiosurgery dose distributions were measured using the radiochromic film in a head phantom undergoing the same treatment as a patient, and were compared with the planned distributions. For an example case (nominal 2.0‐cm‐diam cone), film measurement verified the calculated dose distribution in one plane. The simple measurement technique described led to experimental uncertainties of ±0.1 cm for the 90% and 50% isodose lines, ±0.3 cm for the 20% line, and ±0.5 cm for the 10% line. Isocenter dose was measured with an uncertainty of ±3%. Refinements to the technique should allow more precise measurements. It is concluded that the radiochromic film, with some limitations, is a convenient and useful tool for dynamic radiosurgery quality assurance.