Successful Disinfection of Needleless Access Ports: A Matter of Time and Friction
Authors:
- Wendy KalerWKWendy Kaler
- Raymond ChinnRCRaymond Chinn
- 1 January 2007
- journal article
- Published by Association for Vascular Access in Journal of the Association for Vascular Access
- Vol. 12 (3) , 140-142
Abstract
Background: There is controversy as to whether the design of the needleless access port or the method used to clean it prior to access impacts successful disinfection. The authors studied the disinfection effectiveness for needleless access ports. Method: The ports of 4 models of needleless access ports were inoculated with bacteria. The ports were disinfected for 15 seconds with 70% alcohol alone or 3.15% chlorhexidine/70% alcohol (Chlorascrub, PDI, Orangeburg, NY). Saline flush solutions were collected and cultured. Results: Disinfection with either 70% alcohol alone or with 3.15% chlorhexidine/70% alcohol for 15 seconds was effective. Conclusions: All models of needleless access ports were effectively disinfected using these two methods.Keywords
This publication has 6 references indexed in Scilit:
- Outbreak of Bloodstream Infection Temporally Associated with the Use of an Intravascular Needleless ValveClinical Infectious Diseases, 2007
- Disinfection of Needleless Catheter Connectors and Access Ports With Alcohol May Not Prevent Microbial Entry: The Promise of a Novel Antiseptic-Barrier CapInfection Control & Hospital Epidemiology, 2006
- Increased Catheter-Related Bloodstream Infection Rates After the Introduction of a New Mechanical Valve Intravenous Access PortInfection Control & Hospital Epidemiology, 2006
- Blue Ribbon Abstract AwardAmerican Journal of Infection Control, 2005
- Guidelines for the Prevention of Intravascular Catheter–Related InfectionsInfection Control & Hospital Epidemiology, 2002
- Prospective randomised trial of povidone-iodine, alcohol, and chlorhexidine for prevention of infection associated with central venous and arterial cathetersThe Lancet, 1991