Internal Calibration Potentiometric Aptasensors for Simultaneous Detection of Hg2+, Cd2+, and As3+ Based on a Screen-Printed Carbon Electrodes Array
Authors:
- 25 June 2018
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 90 (14) , 8337-8344
Abstract
An all-solid-state potentiometric aptasensor array based on a multi-channel disposable screen-printed carbon electrode (SPCE) was demonstrated for the simultaneous detection of Hg2+, Cd2+, and As3+ by open circuit potential (OCP) technology. The potential of the channel with an internal calibration DNA sequence (IC-DNA) was employed as the internal calibration potential (ICP) to subtract the background signal generated by the detection system, providing a built-in correction methodology. As a result, the developed aptasensor array showed high sensitivity and accuracy for detecting Hg2+, Cd2+, and As3+ without mutual interference or interference from other ions. The linear response ranged from 2.5 pM to 2.5 μM, and the detection limits for Hg2+, Cd2+, and As3+ were 2.0, 0.62, and 0.17 pM, respectively. Furthermore, the potentiometric aptasensor array was successfully applied for the simultaneous detection of three ions in real samples. The results obtained from the developed approach agreed well with the results obtained from inductively coupled plasma mass spectrometry.Keywords
Funding Information
- National Natural Science Foundation of China (21405049, 21575042)
This publication has 55 references indexed in Scilit:
- Colloidal nanoparticles as advanced biological sensorsScience, 2014
- Rapid speciation analysis of mercury in seawater and marine fish by cation exchange chromatography hyphenated with inductively coupled plasma mass spectrometryJournal of Chromatography A, 2013
- Global Biogeochemical Cycling of Mercury: A ReviewAnnual Review of Environment and Resources, 2009
- High-Throughput Elemental Analysis of Small Aqueous Samples by Emission Spectrometry with a Compact, Atmospheric-Pressure Solution-Cathode Glow DischargeAnalytical Chemistry, 2007
- Low dose mercury toxicity and human healthEnvironmental Toxicology and Pharmacology, 2005
- The application of ICPSMS, GFAAS and HGAFS to the analysis of water and sediment samples from a temperate stratified estuaryJournal of Environmental Monitoring, 2001
- Determination of Mercury in Drinking Water by Inductively Coupled Plasma Atomic Emission Spectrometry with Electrothermal VaporizationApplied Spectroscopy, 1985
- Relative effect of molecular absorption on atomic absorption and atomic fluorescenceAnalytical Chemistry, 1974
- Atomic fluorescence spectrometryAnalytical Chemistry, 1971
- Study of Experimental Parameters in Atomic Fluorescence Flame Spectrometry.Analytical Chemistry, 1964