Computational Study of the Blood Flow in Three Types of 3D Hollow Fiber Membrane Bundles
Open Access
- 6 November 2013
- journal article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 135 (12) , 121009-12100911
Abstract
The goal of this study is to develop a computational fluid dynamics (CFD) modeling approach to better estimate the blood flow dynamics in the bundles of the hollow fiber membrane based medical devices (i.e., blood oxygenators, artificial lungs, and hemodialyzers). Three representative types of arrays, square, diagonal, and random with the porosity value of 0.55, were studied. In addition, a 3D array with the same porosity was studied. The flow fields between the individual fibers in these arrays at selected Reynolds numbers (Re) were simulated with CFD modeling. Hemolysis is not significant in the fiber bundles but the platelet activation may be essential. For each type of array, the average wall shear stress is linearly proportional to the Re. For the same Re but different arrays, the average wall shear stress also exhibits a linear dependency on the pressure difference across arrays, while Darcy's law prescribes a power-law relationship, therefore, underestimating the shear stress level. For the same Re, the average wall shear stress of the diagonal array is approximately 3.1, 1.8, and 2.0 times larger than that of the square, random, and 3D arrays, respectively. A coefficient C is suggested to correlate the CFD predicted data with the analytical solution, and C is 1.16, 1.51, and 2.05 for the square, random, and diagonal arrays in this paper, respectively. It is worth noting that C is strongly dependent on the array geometrical properties, whereas it is weakly dependent on the flow field. Additionally, the 3D fiber bundle simulation results show that the three-dimensional effect is not negligible. Specifically, velocity and shear stress distribution can vary significantly along the fiber axial direction.Keywords
This publication has 38 references indexed in Scilit:
- A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis IndexJournal of Biomechanical Engineering, 2012
- Computational Model‐Based Design of a Wearable Artificial Pump‐Lung for Cardiopulmonary/Respiratory SupportArtificial Organs, 2011
- Micro-scale modeling of flow and oxygen transfer in hollow-fiber membrane bundleJournal of Membrane Science, 2010
- Device Thrombogenicity Emulator (DTE) − Design optimization methodology for cardiovascular devices: A study in two bileaflet MHV designsJournal of Biomechanics, 2010
- Computational Design and In Vitro Characterization of an Integrated Maglev Pump‐OxygenatorArtificial Organs, 2009
- Computational Fluid Flow and Mass Transfer of a Functionally Integrated Pediatric Pump-Oxygenator ConfigurationASAIO Journal, 2008
- Modeling of Blood Flow in a Balloon-Pulsed Intravascular Respiratory CatheterASAIO Journal, 2007
- Characterization of membrane blood oxygenation devices using computational fluid dynamicsJournal of Membrane Science, 2007
- Blood flow and oxygen transfer rate of an outside blood flow membrane oxygenatorJournal of Membrane Science, 2000
- Theoretical and Experimental Evaluation for Blood Pressure Drop and Oxygen Transfer Rate in Outside Blood Flow Membrane Oxygenator.JOURNAL OF CHEMICAL ENGINEERING OF JAPAN, 1999