A simple method for rectified noise floor suppression: Phase‐corrected real data reconstruction with application to diffusion‐weighted imaging
Open Access
- 25 May 2010
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 64 (2) , 418-429
Abstract
Diffusion‐weighted MRI is an intrinsically low signal‐to‐noise ratio application due to the application of diffusion‐weighting gradients and the consequent longer echo times. The signal‐to‐noise ratio worsens with increasing image resolution and diffusion imaging methods that use multiple and higher b‐values. At low signal‐to‐noise ratios, standard magnitude reconstructed diffusion‐weighted images are confounded by the existence of a rectified noise floor, producing poor estimates of diffusion metrics. Herein, we present a simple method of rectified noise floor suppression that involves phase correction of the real data. This approach was evaluated for diffusion‐weighted imaging data, obtained from ethanol and water phantoms and the brain of a healthy volunteer. The parameter fits from monoexponential, biexponential, and stretched‐exponential diffusion models were computed using phase‐corrected real data and magnitude data. The results demonstrate that this newly developed simple approach of using phase‐corrected real images acts to reduce or even suppress the confounding effects of a rectified noise floor, thereby producing more accurate estimates of diffusion parameters. Magn Reson Med, 2010.Keywords
This publication has 28 references indexed in Scilit:
- Using Human and Model Performance to Compare MRI ReconstructionsIEEE Transactions on Medical Imaging, 2006
- Characterization of continuously distributed cortical water diffusion rates with a stretched‐exponential modelMagnetic Resonance in Medicine, 2003
- Biexponential apparent diffusion coefficient parametrization in adult vs newborn brainMagnetic Resonance Imaging, 2001
- Test liquids for quantitative MRI measurements of self-diffusion coefficient in vivoMagnetic Resonance in Medicine, 2000
- Multi-component apparent diffusion coefficients in human brainNMR in Biomedicine, 1999
- Estimation of the Noise in Magnitude MR ImagesMagnetic Resonance Imaging, 1998
- Flow compensation in MRI using a phase‐corrected real reconstructionMagnetic Resonance in Medicine, 1993
- Relaxation in the subcritical pitchfork bifurcation: From critical to Gaussian scalingPhysical Review A, 1991
- Regional T2 and Sodium Concentration Estimates in the Normal Human Brain by Sodium-23 MR Imaging at 1.5 TJournal of Computer Assisted Tomography, 1989
- Nuclear Magnetic Resonance Transverse Relaxation Times of Water Protons in Skeletal MuscleBiophysical Journal, 1974