A new conducting polymer glucose sensor based on polythianaphthene
Authors:
- 1 December 1995
- journal article
- research article
- Published by Wiley in Electroanalysis
- Vol. 7 (12) , 1151-1155
Abstract
A glucose biosensor has been developed by entrapment of glucose oxidase in a polythianaphthene film formed onto a platinum surface. Glucose has been measured by H2O2 detection at +400 mV (vs. SCE). Electrochemical kinetic constants for the enzyme electrode have been calculated and indicated that in the unsaturated region the rate limiting step is the diffusion of the substrate through the polymer. The biosensor exhibited a linear response in the range 0.05–7.4 mmol/L with a response time of 30–60s and a relative standard deviation of 5.4%. The presence of the film onto the electrode prevents the oxidation of interfering agents such as ascorbic and uric acid. The use of additional polycarbonate membranes onto the electrode surface extended the linearity of the response allowing the probe to be used in blood glucose analysis, The amount of glucose in two control sera was measured using the biosensor. Results correlated well with the glucose nominal values.Keywords
This publication has 18 references indexed in Scilit:
- Homogeneous mechanism of ascorbic acid interference in hydrogen peroxide detection at enzyme-modified electrodesAnalytical Chemistry, 1992
- Electrochemical immobilization of enzymes. 3. Immobilization of glucose oxidase in thin films of electrochemically polymerized phenolsAnalytical Chemistry, 1992
- Amperometric polypyrrole enzyme electrodes: effect of permeability and enzyme locationSensors and Actuators B: Chemical, 1992
- Electropolymerized films to prevent interferences and electrode fouling in biosensorsBiosensors and Bioelectronics, 1991
- Direct electron transfer with glucose oxidase immobilized in an electropolymerized poly( N-methylpyrrole) film on a gold microelectrodeAnalytica Chimica Acta, 1990
- Optimization of a polypyrrole glucose oxidase biosensorBiosensors and Bioelectronics, 1990
- Simultaneous immobilization of glucose oxidase and a mediator in conducting polymer filmsJournal of the Chemical Society, Chemical Communications, 1988
- Protein-modified electrodes. The glucose oxidase/polypyrrole systemAnalytical Chemistry, 1986
- Electrically conducting polymers: a review of the electropolymerization reaction, of the effects of chemical structure on polymer film properties, and of applications towards technologyCanadian Journal of Chemistry, 1986
- Photon correlation spectroscopy study on the stability of small unilamellar DPPC vesiclesBiophysical Journal, 1982