The calibration and use of pressure transducers in tensiometer systems
- 1 April 1993
- journal article
- Published by Wiley in Hydrological Processes
- Vol. 7 (2) , 205-211
Abstract
The uncertainty in a transducer/tensiometer system was assessed with temperature and pressure calibrations. A reference transducer/tensiometer pair was used to factor out temperature related deviations from two monitoring pairs. The reference pair removed most of the deviations, resulting in a high estimate of precision. In contrast to earlier reports of high accuracy, these estimates of accuracy were considerably reduced by a time correlated residual pattern. The calibrations suggested that the electronic components may be responsible for these residual errors and illustrated the need for experimentation which isolates the error among groups of components. The complexity of transducer/tensiometer networks, and the differing response of each component to thermal loading, demonstrated the necessity of using a reference system, which when properly designed can yield reliable pressure readings for soil water.Keywords
This publication has 12 references indexed in Scilit:
- Calibration and use of pressure transducers in soil hydrologyHydrological Processes, 1989
- Instrumentation System for Determining Watershed Hydrologic CharacteristicsTransactions of the ASAE, 1987
- An Electrical Readout System for TensiometersSoil Science Society of America Journal, 1986
- ERRORS IN READING TENSIOMETER VACUA WITH PRESSURE TRANSDUCERSSoil Science, 1984
- An Electronic Transducer for Continuous Water Level MonitoringTransactions of the ASAE, 1984
- A Pressure Transducer for Field TensiometersSoil Science Society of America Journal, 1983
- AN AUTOMATED SYSTEM FOR MEASURING SOIL WATER POTENTIAL GRADIENTS IN A RHIZOTRON SOIL PROFILESoil Science, 1980
- Automatic monitoring of soil moisture conditions in a hillslope spur and hollowJournal of Hydrology, 1977
- Diurnal and Seasonal Soil Water Uptake and Flux within a Bermudagrass Root ZoneSoil Science Society of America Journal, 1975
- Tensiometer-Pressure Transducer System for Studying Unsteady Flow Through SoilsTransactions of the ASAE, 1972