Estimation of Hounsfield unit conversion parameters for pelvic CT images
- 7 June 2018
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 41 (3) , 739-745
Abstract
Linear scaling is used to convert raw computed tomography (CT) pixel values into Hounsfield units corresponding to different tissue values. Analysis of a benchmarking study, presented here, where the same CT scan was imported into and then exported from multiple radiotherapy treatment planning systems, found inconsistencies in HU scaling parameter values exported along with the images, particularly when images were exported using the Radiation Therapy Oncology Group format. Several methods of estimating conversion parameters, based on estimating pixel values corresponding to air and water within the image, for pelvic CT images from a large multi-centre trial were compared against original Digital Imaging and Communications in Medicine export parameters. In general using the mean of a sample region at the centroid of the bladder to estimate the value of water was more accurate than using the minimum or maximum or a single value at the centroid. Accuracy of methods of air estimation tested were dependent in part on features of the CT scanners and treatment planning systems, making it difficult to pick one method as superior that was independent of scanner and treatment planning system type. Based on the above analysis, methods for estimating air and water were selected for use in performing linear scaling of a set of pelvic CT images prior to their use in an interpatient image registration application. The selected methods were validated against a more recent and homogeneous dataset. Estimation error was found to be much lower within the validation set.Keywords
Funding Information
- National Health and Medical Research Council (1077788)
This publication has 11 references indexed in Scilit:
- Non-traumatic bladder rupture showing less than 10 Hounsfield units of ascitesAcute Medicine & Surgery, 2016
- Automatic Substitute Computed Tomography Generation and Contouring for Magnetic Resonance Imaging (MRI)-Alone External Beam Radiation Therapy From Standard MRI SequencesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2015
- The Characterization of Feces and Urine: A Review of the Literature to Inform Advanced Treatment TechnologyCritical Reviews in Environmental Science and Technology, 2015
- Multi-Atlas-Based Segmentation of Pelvic Structures from CT Scans for Planning in Prostate Cancer Radiotherapy2013
- Effect of androgen deprivation therapy on muscle attenuation in men with prostate cancerJournal of Medical Imaging and Radiation Oncology, 2013
- ‘Rapid Learning health care in oncology’ – An approach towards decision support systems enabling customised radiotherapy’Radiotherapy and Oncology, 2013
- Quality improvements in prostate radiotherapy: Outcomes and impact of comprehensive quality assurance during the TROG 03.04 ‘RADAR’ trialJournal of Medical Imaging and Radiation Oncology, 2013
- Radiomics: Extracting more information from medical images using advanced feature analysisEuropean Journal of Cancer, 2012
- Physiology of Renal Medullary Tip Hyperattenuation at Unenhanced CT: Urinary Specific Gravity and the NaCl Concentration GradientRadiology, 2008
- Detailed review and analysis of complex radiotherapy clinical trial planning data: Evaluation and initial experience with the SWAN software systemRadiotherapy and Oncology, 2008