Electrical conductivity images of biological tissue phantoms in MREIT

Authors:
  • BLByung Il Lee
    Byung Il Lee
  • EWEung Je Woo
    Eung Je Woo
  • SLSoo Yeol Lee
    Soo Yeol Lee
  • TKTae-Seong Kim
    Tae-Seong Kim
  • OKOhin Kwon
    Ohin Kwon
  • JSJin Keun Seo
    Jin Keun Seo

Abstract
We present cross-sectional conductivity images of two biological tissue phantoms. Each of the cylindrical phantoms with both diameter and height of 140 mm contained chunks of biological tissues such as bovine tongue and liver, porcine muscle and chicken breast within a conductive agar gelatin as the background medium. We attached four recessed electrodes on the sides of the phantom with equal spacing among them. Injecting current pulses of 480 or 120 mA ms into the phantom along two different directions, we measured the z-component Bz of the induced magnetic flux density B=(Bx, By, Bz) with a magnetic resonance electrical impedance tomography (MREIT) system based on a 3.0 T MRI scanner. Using the harmonic Bz algorithm, we reconstructed cross-sectional conductivity images from the measured Bz data. Reconstructed images clearly distinguish different tissues in terms of both their shapes and conductivity values. In this paper, we experimentally demonstrate the feasibility of the MREIT technique in producing conductivity images of different biological soft tissues with a high spatial resolution and accuracy when we use a sufficient amount of the injection current.