Complete Loss of ICD Programmability After Magnetic Resonance Imaging
Authors:
- 22 July 2004
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 27 (7) , 1002-1004
Abstract
The purpose of this case report is to describe the effects of an MRI performed on a patient without realizing that an ICD has been previously implanted. After a few seconds of imaging the adversity was recognized and the examination was stopped immediately. The patient was not pacemaker dependent and had neither physical complaints nor electrocardiographic changes in the surface ECG. A consecutively performed ICD assessment showed a backup mode with standard parameters for pacing (VVI 50 beats/min) and arrhythmia detection and treatment. The device could not be programmed by the external programmer. With the exception of printing out the parameters, all software functions were no longer feasible. A device examination by the manufacturer after ICD replacement showed that a major portion of the device memory was corrupt. Even ICDs of a newer generation are susceptible to magnetic interference, with the danger of complete loss of programmability. (PACE 2004; 27:1002–1004)Keywords
This publication has 11 references indexed in Scilit:
- Implantable Cardioverter Defibrillator Dysfunction During and After Magnetic Resonance ImagingPacing and Clinical Electrophysiology, 2002
- Interference with Cardiac Pacemakers by Magnetic Resonance Imaging: Are There Irreversible Changes at 0.5 Tesla?Pacing and Clinical Electrophysiology, 2001
- Force and Torque Effects of a 1.5‐Tesla MRI Scanner on Cardiac Pacemakers and ICDsPacing and Clinical Electrophysiology, 2001
- Pacing in magnetic resonance imaging environment: Clinical and technical considerations on compatibilityEuropean Heart Journal, 2001
- Effects of magnetic resonance imaging on cardiac pacemakers and electrodesAmerican Heart Journal, 1997
- Safe Performance of Magnetic Resonance Imaging on Five Patients with Permanent Cardiac PacemakersPacing and Clinical Electrophysiology, 1996
- Effects of Nuclear Magnetic Resonance Imaging on Cardiac PacemakersPacing and Clinical Electrophysiology, 1995
- ABSTRACT SESSION 22: DEVELOPMENTS IN LEADS AND PULSE GENERATORSPacing and Clinical Electrophysiology, 1994
- Case Report: Nuclear Magnetic Resonance Imaging in a Patient with a PacemakerThe American Journal of the Medical Sciences, 1993
- Nuclear Magnetic Resonance Imaging in a Patient with a DDD PacemakerPacing and Clinical Electrophysiology, 1989