A closed-loop pump-driven wire-guided flow jet for ultrafast spectroscopy of liquid samples
Authors:
- 1 September 2015
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 86 (9) , 093105
Abstract
We describe the design and provide the results of the full characterization of a closed-loop pump-driven wire-guided flow jet system. The jet has excellent optical quality with a wide range of liquids spanning from alcohol to water based solutions, including phosphate buffers used for biological samples. The thickness of the jet film varies depending on the flow rate between 90 μm and 370 μm. The liquid film is very stable, and its thickness varies only by 0.76% under optimal conditions. Measured transmitted signal reveals a long term optical stability (hours) with a RMS of 0.8%, less than the overall noise of the spectroscopy setup used in our experiments. The closed loop nature of the overall jet design has been optimized for the study of precious biological samples, in limited volumes, to remove window contributions from spectroscopic observables. This feature is particularly important for femtosecond studies in the UV range.Keywords
Funding Information
- International Max Planck Research School for Ultrafast Imaging and Structural Dynamics
This publication has 5 references indexed in Scilit:
- A closed-loop pump-driven wire-guided flow jet for ultrafast spectroscopy of liquid samplesReview of Scientific Instruments, 2015
- Liquid-sheet jets for terahertz spectroscopyOptics Express, 2014
- Two-Dimensional Fourier Transform Spectroscopy of Adenine and Uracil Using Shaped Ultrafast Laser Pulses in the Deep UVThe Journal of Physical Chemistry A, 2011
- Flowing liquid sample jet for resonance Raman and ultrafast optical spectroscopyReview of Scientific Instruments, 2003
- Resolution of absorption spectra of rhodamine 6G aggregates in aqueous solution using the law of mass actionChemical Physics Letters, 1997