Static and moving phantom studies for radiation treatment planning in a positron emission tomography and computed tomography (PET/CT) system
- 29 August 2008
- journal article
- Published by Springer Nature in Annals of Nuclear Medicine
- Vol. 22 (7) , 579-586
Abstract
To determine an appropriate threshold value for delineation of the target in positron emission tomography (PET) and to investigate whether PET can delineate an internal target volume (ITV), a series of phantom studies were performed. An ellipse phantom (background) was filled with 1028 Bq/ml of [18F] fluoro-2-deoxyglucose (18FDG), and six spheres of 10 mm, 13 mm, 17 mm, 22 mm, 28 mm, and 37 mm in diameter inside it were filled with 18FDG activity to achieve source-to-background (S/B) ratios of 10, 15, and 20. In static phantom experiments, an appropriate threshold value was determined so that the size of PET delineation fits to an actual sphere. In moving phantom experiments with total translations of 10 mm, 20 mm, and 30 mm and a period of oscillation of 4 s, the maximum size of PET delineation with the appropriate threshold value was measured in both the axial and sagittal planes. In the static phantom experiments, the measured maximum 18FDG activities of spheres of less than 22 mm were lower than 80% of the injected 18FDG activity, and those for the larger spheres ranged from 90% to 110%. Appropriate threshold values determined for the spheres of 22 mm or more ranged from 30% to 40% of the maximum 18FDG activity, independent of the S/B ratio. Therefore, we adopted an appropriate threshold value as 35% of the measured maximum 18FDG activity. In moving phantom experiments, the maximum 18FDG activity of spheres decreased significantly, dependent on the movement distance. Although the sizes of PET delineation with 35% threshold value tended to be slightly smaller (18FDG activity was adopted, the sizes of PET delineation were almost the same for static and moving phantom spheres of 22 mm or more in the axial plane. In addition, PET images have the potential to provide an individualized ITV.Keywords
This publication has 35 references indexed in Scilit:
- Correlation of PET standard uptake value and CT window-level thresholds for target delineation in CT-based radiation treatment planningInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Positron emission tomography/computed tomography: diagnostic accuracy in lymphomaBritish Journal of Haematology, 2006
- Phased attenuation correction in respiration correlated computed tomography/positron emitted tomographyMedical Physics, 2006
- Thresholding in PET images of static and moving targetsPhysics in Medicine & Biology, 2005
- The contribution of integrated PET/CT to the evolving definition of treatment volumes in radiation treatment planning in lung cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- [18F]Fluorodeoxyglucose Uptake by Positron Emission Tomography Predicts Outcome of Non–Small-Cell Lung CancerJournal of Clinical Oncology, 2005
- Is there a role for FGD-PET in radiotherapy planning in esophageal carcinoma?Radiotherapy and Oncology, 2004
- Accuracy of Whole-Body Dual-Modality Fluorine-18–2-Fluoro-2-Deoxy-d-Glucose Positron Emission Tomography and Computed Tomography (FDG-PET/CT) for Tumor Staging in Solid Tumors: Comparison With CT and PETJournal of Clinical Oncology, 2004
- Positron Emission Tomography in Limited-Stage Small-Cell Lung Cancer: A Prospective StudyJournal of Clinical Oncology, 2004
- Randomized Controlled Trial of the Role of Positron Emission Tomography in the Management of Stage I and II Non-Small-Cell Lung CancerJournal of Clinical Oncology, 2004