The first clinical implementation of electromagnetic transponder‐guided MLC tracking
Open Access
- 23 January 2014
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 41 (2) , 020702
Abstract
Purpose:: We report on the clinical process, quality assurance, and geometric and dosimetric results of the first clinical implementation of electromagnetic transponder‐guided MLC tracking which occurred on 28 November 2013 at the Northern Sydney Cancer Centre.Methods:: An electromagnetic transponder‐based positioning system (Calypso) was modified to send the target position output to in‐house‐developed MLC tracking code, which adjusts the leaf positions to optimally align the treatment beam with the real‐time target position. Clinical process and quality assurance procedures were developed and performed. The first clinical implementation of electromagnetic transponder‐guided MLC tracking was for a prostate cancer patient being treated with dual‐arc VMAT (RapidArc). For the first fraction of the first patient treatment of electromagnetic transponder‐guided MLC tracking we recorded the in‐room time and transponder positions, and performed dose reconstruction to estimate the delivered dose and also the dose received had MLC tracking not been used.Results:: The total in‐room time was 21 min with 2 min of beam delivery. No additional time was needed for MLC tracking and there were no beam holds. The average prostate position from the initial setup was 1.2 mm, mostly an anterior shift. Dose reconstruction analysis of the delivered dose with MLC tracking showed similar isodose and target dose volume histograms to the planned treatment and a 4.6% increase in the fractional rectal V60. Dose reconstruction without motion compensation showed a 30% increase in the fractional rectal V60 from that planned, even for the small motion.Conclusions:: The real‐time beam‐target correction method, electromagnetic transponder‐guided MLC tracking, has been translated to the clinic. This achievement represents a milestone in improving geometric and dosimetric accuracy, and by inference treatment outcomes, in cancer radiotherapy.Keywords
This publication has 20 references indexed in Scilit:
- Proof of concept of MRI-guided tracked radiation delivery: tracking one-dimensional motionPhysics in Medicine & Biology, 2012
- Megavoltage Image-Based Dynamic Multileaf Collimator Tracking of a NiTi Stent in Porcine Lungs on a Linear AcceleratorInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Electromagnetic-Guided Dynamic Multileaf Collimator Tracking Enables Motion Management for Intensity-Modulated Arc TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Failure mode and effect analysis‐based quality assurance for dynamic MLC tracking systemsMedical Physics, 2010
- Real‐time tumor tracking: Automatic compensation of target motion using the Siemens 160 MLCMedical Physics, 2010
- Integration of Real-Time Internal Electromagnetic Position Monitoring Coupled With Dynamic Multileaf Collimator Tracking: An Intensity-Modulated Radiation Therapy Feasibility StudyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Toward Submillimeter Accuracy in the Management of Intrafraction Motion: The Integration of Real-Time Internal Position Monitoring and Multileaf Collimator Target TrackingInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Management of three‐dimensional intrafraction motion through real‐time DMLC trackingMedical Physics, 2008
- Geometric accuracy of a real-time target tracking system with dynamic multileaf collimator tracking systemInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Motion adaptive x-ray therapy: a feasibility studyPhysics in Medicine & Biology, 2000