Fast Measurement of Ion Temperature Using Filter Charge Exchange Recombination Spectroscopy System Towards Real-Time Plasma Control in JT-60U
Authors:
- Jt-60 TeamJTJt-60 Team
- 1 February 2007
- journal article
- research article
- Published by Taylor & Francis in Fusion Science and Technology
- Vol. 51 (2T) , 301-303
Abstract
A filter charge exchange recombination spectroscopy (CXRS) system has been developed for the real-time measurement and feedback control of ion temperature (Ti) profiles in JT-60U. In order to improve the dynamic range and the spatial coverage, the passband of each filter used for the fast CXRS diagnostic is optimized for the target value of Ti (˜0.1-20 keV), and the number of relevant channels is increased (maximum: 12 channels). Real-time processor system has also been improved using advantage of the filter CXRS method. Utilizing this filter CXRS system, real-time control of the central Ti has been demonstrated.Keywords
This publication has 6 references indexed in Scilit:
- Observations of the collapse and recovery of the temperature pedestal using diagnostics with the fast temporal resolution in JT-60UPlasma Physics and Controlled Fusion, 2006
- Development of Fast Charge Exchange Recombination Spectroscopy by Using Interference Filter Method in JT-60UJournal of Plasma and Fusion Research, 2003
- Advanced Real-Time Feedback Control in JT-60U High Performance Discharges for Application to Fusion Reactor PlasmasFusion Science and Technology, 2002
- Multichordal charge exchange recombination spectroscopy on the JT-60U tokamakReview of Scientific Instruments, 2001
- Electron cyclotron emission measurements in JT-60UFusion Engineering and Design, 2001
- Observations on the formation and control of transport barriersPlasma Physics and Controlled Fusion, 2000