A low‐flow to‐and‐fro system
Authors:
Open Access
- 1 November 1991
- journal article
- Published by Wiley in Anaesthesia
- Vol. 46 (11) , 948-951
Abstract
Summary: The insertion of a long deadspace tube between an anaesthetic breathing system and a ventilator produces only imperfect separation of the breathing system gas from the driving gas. This laboratory study has investigated different connecting tubes to establish the maximum tidal volume possible before ventilator gas contaminates the gas in a low‐flow to‐and‐fro system (fresh gas flow 1 litre/minute). A larger volume tube enables the use of larger tidal volumes, and plastic corrugated tubes are slightly better than black rubber corrugated tubes in this respect. The maximum tidal volume possible without contamination decreases as ventilatory rate increases, but the maximum minute volume is increased. A 22 mm plastic corrugated tube of internal volume 1.5 litres should be adequate for clinical use with the to‐and‐fro system described in this study at afresh gas flow of 1 litre/minute.Keywords
This publication has 9 references indexed in Scilit:
- A low flow open circle system for anaesthesia Part I: Laboratory evaluation at normal and high frequenciesAnaesthesia, 1990
- Evaluation of a multigas anaesthetic monitor: the Datex CapnomacAnaesthesia, 1988
- Relationship between frequency of ventilation, airways and pulmonary artery pressures, cardiac output and tracheal tube deadspaceActa Anaesthesiologica Scandinavica, 1986
- A NEW VALVELESS ALL-PURPOSE VENTILATORBritish Journal of Anaesthesia, 1983
- A METHOD OF CONTROLLED VENTILATION WITH A CIRCLE-ABSORBER BREATHING SYSTEMBritish Journal of Anaesthesia, 1977
- CONTROLLED VENTILATION WITH THE MAPLESON D SYSTEM: A Theoretical and Experimental StudyBritish Journal of Anaesthesia, 1977
- The Bain circuitAnaesthesia, 1977
- Evaluation of the lung ventilator performance analyser*Anaesthesia, 1976
- Ventilation Standards for Use in Artificial RespirationJournal of Applied Physiology, 1955