Coronary CT angiography in obese patients using 3rd generation dual-source CT: effect of body mass index on image quality
- 28 December 2015
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 26 (9) , 2937-2946
Abstract
To evaluate the image quality of coronary CT angiography (CCTA) in obese patients using a 3rd generation, dual-source CT scanner. We retrospectively evaluated 102 overweight and obese patients who had undergone CCTA. Studies were performed with 3rd generation dual-source CT, prospectively ECG-triggered acquisition at 120 kV, and automated tube current modulation. Advanced modeled iterative reconstruction was used. Patients were divided into three BMI groups: 1)25–29.9 kg/m2; 2)30–39.9 kg/m2; 3) ≥ 40 kg/m2. Vascular attenuation in the coronary arteries was measured. Contrast-to-noise ratio (CNR) was calculated. Image quality was subjectively evaluated using five-point scales. Image quality was considered diagnostic in 97.6 % of examinations. CNR was consistently adequate in all groups but decreased for groups 2 and 3 in comparison to group 1 as well as for group 3 compared to group 2 (p = 0.001, respectively). Subjective image quality was significantly higher in group 1 compared to group 3 (attenuation proximal: 4.8 ± 0.4 vs. 4.4 ± 0.6, p = 0.011; attenuation distal: 4.5 ± 0.7 vs. 4.0 ± 0.8, p = 0.019; noise: 4.7 ± 0.6 vs. 3.8 ± 0.7, p < 0.001). The mean effective dose was 9.5 ± 3.9 mSv for group 1, 11.4 ± 4.7 mSv for group 2 and 14.0 ± 6.4 mSv for group 3. Diagnostic image quality can be routinely obtained at CCTA in obese patients with 3rd generation DSCT at 120 kV. • Diagnostic CCTA can be routinely performed in obese patients with 3 rd generation DSCT. • 120-kV tube voltage allows diagnostic image quality in patients with BMI > 40 kg/m 2 . • 80-ml contrast medium can be administered without significant decline in vascular attenuation.Keywords
This publication has 30 references indexed in Scilit:
- Prospectively ECG-triggered high-pitch coronary angiography with third-generation dual-source CT at 70 kVp tube voltage: Feasibility, image quality, radiation dose, and effect of iterative reconstructionJournal of Cardiovascular Computed Tomography, 2014
- Iterative model reconstruction: Improved image quality of low-tube-voltage prospective ECG-gated coronary CT angiography images at 256-slice CTEuropean Journal of Radiology, 2014
- Accuracy of prospectively ECG-triggered very low-dose coronary dual-source CT angiography using iterative reconstruction for the detection of coronary artery stenosis: comparison with invasive catheterizationEuropean Heart Journal - Cardiovascular Imaging, 2014
- Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom studyEuropean Radiology, 2014
- Iterative Reconstruction to Preserve Image Quality and Diagnostic Accuracy at Reduced Radiation Dose in Coronary CT AngiographyJACC: Cardiovascular Imaging, 2013
- Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery diseaseEuropean Radiology, 2013
- Image quality and radiation dose of low dose coronary CT angiography in obese patients: Sinogram affirmed iterative reconstruction versus filtered back projectionEuropean Journal of Radiology, 2012
- Effect of Increasing Body Mass Index on Image Quality and Positive Predictive Value of 100-kV Coronary Computed Tomographic AngiographyThe American Journal of Cardiology, 2010
- Intravenous Contrast Medium Administration and Scan Timing at CT: Considerations and ApproachesRadiology, 2010
- Multidetector Computed Tomography for the Diagnosis of Coronary Artery Disease: A Systematic ReviewThe American Journal of Medicine, 2006