Impact of mitral valve intervention with left ventricular assist device implantation on postoperative outcomes and morphologic change
Authors:
- 27 November 2017
- journal article
- research article
- Published by Springer Nature in Journal of Artificial Organs
- Vol. 21 (2) , 164-171
Abstract
Although mitral regurgitation (MR) is prevalent in patients with end-stage heart failure, the impact of mitral valve (MV) surgery on outcomes after left ventricular assist device (LVAD) implantation and morphologic changes of MV remains unclear. We retrospectively reviewed 74 patients who underwent LVAD implantation as a bridge to transplant. Of these, 11 (15%) underwent MV repair concomitant with or prior to LVAD implantation, while 27 patients with preoperative significant (moderate or greater) MR did not undergo concomitant MV surgery. The mean interval between LVAD implantation and the last echocardiographic examination was 913 days. Irrespective of MV surgery, significant LV reverse remodeling including decreased LV and left atrial dimension and improved MR severity was observed in all patients except for patients with prior MV surgery. Histopathological examination of explanted hearts removed at heart transplantation (n = 69) or autopsy (n = 5) revealed that the MV annulus was still dilated (mean perimeter 11.7 cm) in the patients with preoperative significant MR and no concomitant MV surgery. Concomitant MV surgery at the time of LVAD implantation for significant MR might not be always necessary for bridge to transplant or destination therapy cases. However, it might be required in patients having potential for cardiac recovery or patients with severe pulmonary hypertension and depressed right ventricle.Keywords
Funding Information
- Japan Society for the Promotion of Science (JP16K10446)
This publication has 32 references indexed in Scilit:
- Mitral Valve Repair at the Time of Continuous-Flow Left Ventricular Assist Device Implantation Confers Meaningful Decrement in Pulmonary Vascular ResistanceASAIO Journal, 2013
- The 2013 International Society for Heart and Lung Transplantation Guidelines for mechanical circulatory support: Executive summaryThe Journal of Heart and Lung Transplantation, 2013
- Left Ventricular Reverse Remodeling With a Continuous Flow Left Ventricular Assist Device Measured by Left Ventricular End-Diastolic Dimensions and Severity of Mitral RegurgitationASAIO Journal, 2012
- Late Aortic Insufficiency Related to Poor Prognosis During Left Ventricular Assist Device SupportThe Annals of Thoracic Surgery, 2011
- Echocardiographic Findings in Stable Outpatients with Properly Functioning HeartMate II Left Ventricular Assist DevicesJournal of the American Society of Echocardiography, 2011
- Clinical management of continuous-flow left ventricular assist devices in advanced heart failureThe Journal of Heart and Lung Transplantation, 2010
- Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist DeviceThe New England journal of medicine, 2009
- Low Operative Mortality With Implantation of a Continuous-Flow Left Ventricular Assist Device and Impact of Concurrent Cardiac ProceduresCirculation, 2009
- Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist DeviceJACC, 2009
- Use of a Continuous-Flow Device in Patients Awaiting Heart TransplantationNew England Journal of Medicine, 2007