Radiation dose reduction to breast and thyroid during MDCT: effectiveness of an in-plane bismuth shield
Authors:
- 1 July 2006
- journal article
- research article
- Published by SAGE Publications in Acta Radiologica
- Vol. 47 (6) , 562-567
Abstract
Purpose: To evaluate dose reduction and image deterioration using in-plane bismuth breast-shielding and thyroid-shielding for MDCT. Material and Methods: Skin and organ doses of thyroid and breast were measured with thermoluminescent dosimeters in a female Alderson-Rando Phantom with and without a 4-ply in-plane bismuth shield. Routine neck (120 kVp, 150 mAseff; 16×1.5 mm) and chest (120 kVp, 100 mAseff; 16×1.5 mm) scan protocols were simulated on a 16-row MDCT scanner in three different settings: without shielding, with the shield directly on the surface, and with a 1-cm-thick cotton spacer between surface and shield. Image noise was quantified and compared using the t test. Results: On average, shielding resulted in a 47% organ-dose reduction for the thyroid and 32% for the breast. Placement of the spacer between shield and surface had no significant impact on the measured doses, but significantly decreased the image noise ( P<0.05). Conclusion: In-plane bismuth breast and thyroid shielding significantly decreases radiation dose in MDCT without deteriorating image quality.Keywords
This publication has 7 references indexed in Scilit:
- Radiation Dose Reduction to the Male Gonads During MDCT: The Effectiveness of a Lead ShieldAmerican Journal of Roentgenology, 2005
- In-Plane Bismuth Breast Shields for Pediatric CT: Effects on Radiation Dose and Image Quality Using Experimental and Clinical DataAmerican Journal of Roentgenology, 2003
- Orbital, thyroid, and breast superficial radiation shielding for patients undergoing diagnostic CTSeminars in Ultrasound, CT and MRI, 2002
- Radioprotection to the eye during CT scanningAmerican Journal of Ophthalmology, 2001
- Estimated Risks of Radiation-Induced Fatal Cancer from Pediatric CTAmerican Journal of Roentgenology, 2001
- A PC program for estimating organ dose and effective dose values in computed tomographyEuropean Radiology, 1999
- The breast: in-plane x-ray protection during diagnostic thoracic CT--shielding with bismuth radioprotective garments.Radiology, 1997