Abstract
Summary: Background:  We have shown previously that children with developmental disabilities have three times higher incidence of sedation‐related hypoxia when compared with normal children.Objectives:  Our objectives were to describe the changes in endtidal carbon dioxide (ETCO2) values and the utility of ETCO2 monitoring in earlier identification of hypoxia during sedation for brain magnetic resonance imaging (MRI) in children with developmental disabilities.Methods:  We conducted a prospective observational study of a convenience sample of 150 children with developmental disabilities aged 1–10 years who received intravenous sedation for brain MRI. Children were sedated and monitored according to the institution’s sedation protocol. We recorded ETCO2 levels, hypoxia, and adverse events during sedation. Hypoxia was defined as SpO2 < 93%. A change in ETCO2 level ≥ 10 mm Hg from presedation baseline, an intra‐sedation ≥ 50 mm Hg, and loss of capnographic waveform were considered as significant ETCO2 abnormalities.Results:  Of the children, 80.7% (121/150) were sedated with a combination of pentobarbital and fentanyl. ETCO2 abnormalities were noted in 42.6% (64/150) of sedation encounters. Hypoxia occurred in 18% (27/150) of subjects. ETCO2 abnormalities were documented in 19(70%) patients with hypoxia before changes in pulse oximetry were noted. ETCO2 changes were noted a mean of 4.38 ± 1.89 min prior to occurrence of hypoxia.Conclusions:  ETCO2 abnormalities and hypoxia occur commonly during sedation in children with developmental disabilities. ETCO2 monitoring is useful in early recognition of impending hypoxia during sedation in children with developmental disabilities.

This publication has 19 references indexed in Scilit: