The effect of MR surface coils on PET quantification in whole-body PET/MR: Results from a pseudo-PET/MR phantom study
Authors:
- 9 May 2011
- journal article
- nuclear medicine-physics
- Published by Wiley in Medical Physics
- Vol. 38 (5) , 2795-2805
Abstract
Purpose: The use of magnetic resonance (MR) radiofrequency (RF) surface coils is a prerequisite for high‐quality positron emission tomography (PET)/MR imaging. In lack of in‐gantry transmission (TX) sources, the exact position of the RF coils is unknown in PET/MR, and may, therefore, lead to false attenuation correction (AC) of the emission (EM) data. The authors assess lesion and background quantification in AC‐PET by mimicking different PET/MR imaging situations using a whole‐body (WB) PET‐only tomograph. Methods: Phantom experiments were performed on a PET tomograph with68 Ge‐rod TX sources. First, a 15‐cm plastic cylinder was filled uniformly with [18F]‐FDG to simulate a head study. Second, a NEMA NU‐2001 image quality phantom (35 × 25 × 25 cm3) was filled uniformly with [18F]‐FDG to simulate torso imaging. The phantom contained six lesions (10–38 mm diameter, lesion‐to‐background ratio 6:1) centred around a 5 cm diameter lung insert. EM and TX measurements were acquired with and without MR head (cylinder) and surface (NU‐2001 phantom) RF coils in place. The following imaging situations were mimicked in both head and torso phantom studies: (1) PET scan without MR coils in EM and TX for reference, (2) PET scan with coils in both EM and TX, and (3) PET scan with coils in EM but without coils in TX. Two more set‐ups were performed for the torso phantom: (4) PET scan with coils in EM only and phantom shifted slightly compared to (3), and (5) PET scan with coils in EM and TX following local displacement of the surface coils. PET EM data (1)–(4) were corrected for attenuation and scatter using cold TX data. Imaging situations (1)–(3) were repeated with the cylinder phantom and head coil in a combined PET/MR prototype system employing template‐based AC. Results: Head phantom: In case the MR head coils were not accounted for during AC (3), central and peripheral background activity concentration was underestimated by 13%–19% when compared to the reference setup (1). The effects of MR coil omission during AC was replicated in the repeat study with the combined PET/MR prototype. Torso phantom: All lesions were equally visible on all AC‐PET images. The effects of disregarding MR surface RF coils during AC [(3) vs (1)] were 4%, or less. A slightly higher bias was observed when accounting for the RF surface coils that were shifted between EM and TX (5). The effect of coil misalignment and neglect during AC on the quantification of the simulated lungs was insignificant compared to the noise levels in AC‐PET. Conclusions: Unaccounted attenuation from MR surface coils causes a regional bias of AC‐PET data in body regions near the MR coils. Bias of central regions was more noticeable in smaller‐size objects. In torso studies with body surface coils, the visibility of central lesions on PET was unaffected by MR coils following incomplete AC. Coil misalignment of several cm between emission and attenuation images causes an error that was comparable to that arising from unaccounted MR coil attenuation but small compared to the average standard deviation of the activity concentration levels.Keywords
This publication has 23 references indexed in Scilit:
- Toward Implementing an MRI-Based PET Attenuation-Correction Method for Neurologic Studies on the MR-PET Brain PrototypeJournal of Nuclear Medicine, 2010
- Evaluation of the attenuation properties of MR equipment for its use in a whole-body PET/MR scannerPhysics in Medicine & Biology, 2010
- MRI-Based Attenuation Correction for PET/MRI Using Ultrashort Echo Time SequencesJournal of Nuclear Medicine, 2010
- PET/MRI: Paving the Way for the Next Generation of Clinical Multimodality Imaging ApplicationsJournal of Nuclear Medicine, 2010
- Combined 18F-FDG and Fluoride Approach in PET/CT Imaging: Is There a Clinical Future?Journal of Nuclear Medicine, 2009
- FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0European Journal of Nuclear Medicine and Molecular Imaging, 2009
- Tissue Classification as a Potential Approach for Attenuation Correction in Whole-Body PET/MRI: Evaluation with PET/CT DataJournal of Nuclear Medicine, 2009
- MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas RegistrationJournal of Nuclear Medicine, 2008
- Multimodality imaging of structure and functionPhysics in Medicine & Biology, 2008
- Performance evaluation of the whole-body PET scanner ECAT EXACT HR/sup +/ following the IEC standardIEEE Transactions on Nuclear Science, 1997