Parametrization of mammography normalized average glandular dose tables
Authors:
- Xizeng WuXWXizeng Wu
- 1 April 1997
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 24 (4) , 547-554
Abstract
Data from existing tables of normalized glandular doses in mammography were parametrized to determine analytical expressions that match tabulated results within known uncertainties. The parametrization was performed for three different target/filter combinations (molybdenum target–molybdenum filter, molybdenum target–rhodium filter, and rhodium target–rhodium filter) and three different breast compositions (100% adipose, 50% adipose–50% glandular, and 100% glandular). The analytical expressions provide the normalized glandular dose for any breast composition within stated ranges of tabulated input parameters (kVp, half‐value layer, and breast thickness). The maximum difference between tabulated and parametrized data is 1.29%, which is smaller than the stated 2% uncertainty in tabulated dose data due to uncertainties in the x‐ray spectra models used to generate the tables. Analytical expressions are easily coded to create custom functions that return the normalized glandular dose for the set of input parameters. Examples of implementation are presented in Microsoft Visual Basic for Applications™ (VBA).Keywords