Interference in the autoDELFIA ® hAFP immunoassay and effect on second-trimester Down's syndrome screening

Authors:

Abstract
Background: Falsely decreased serum alphafetoprotein (AFP) concentrations are reported in the autoDELFIA ® hAFP immunoassay due to interference by complement. AFP is measured, using this assay, as part of second-trimester and integrated Down's syndrome screening tests. Decreased AFP concentrations increase the calculated risk of Down's syndrome; therefore falsely low AFP, due to assay interference, may artificially increase a patient's risk, and have the potential to cause false screen positive results. It was our aim to assess whether negative interference in the autoDELFIA ® hAFP assay was a cause of very low AFP concentrations, and to examine the effect of falsely decreased concentrations on the calculated risk of Down's syndrome. Methods: Three hundred and twenty-three sequential Down's screening serum samples with very low serum AFP concentration (® hAFP immunoassay were selected and AFP re-measured using the E170 AFP immunoassay. Results: Interference was detected in nine samples (from eight patients) on the basis of discordant AFP concentrations. The interference decreased following storage of samples at 4°C to deplete complement. Use of the falsely low AFP concentrations to calculate risk of Down's syndrome resulted in significantly increased calculated risk compared with complement depleted results. Conclusions: Laboratories should be aware that falsely low AFP concentrations due to complement interference may be obtained using the autoDELFIA ® hAFP immunoassay. We have shown that falsely low AFP concentrations increase the calculated risk of Down's syndrome. This is a potential cause of false Down's syndrome screen positive results.

This publication has 12 references indexed in Scilit: