Abstract
We have previously demonstrated that accelerometer-based vibration analysis detects thromboembolism and pump thrombosis in HeartWare Left Ventricular Assist Device (HVAD) using the third harmonic frequency (pump_speedx3). Thromboembolism also affected the amplitude of the nonharmonic frequencies. The aim of this study was to determine whether nonharmonic-amplitude (NHA) analysis can improve the diagnosis of thromboembolic complications. An accelerometer was attached to HVAD in three in vitro and seven in vivo experiments. Control interventions, including load and pump speed alternations (n = 107), were followed by thromboembolic events (n = 60). A sliding fast-Fourier-transform was analyzed, and changes in NHAs were quantified in the acute phase and in a steady state. Receiver operating characteristic curves were constructed with cutoff values of NHA to detect thromboembolic events. Positive predictive values were calculated on the basis of a specificity of 1. In the acute phase, NHA change was 6.5 times higher under thromboembolism than under control interventions (p < 0.001). Most thromboembolic events lead to concomitant changes in both NHA and third-harmonic amplitude. Combining the two methods improved the PPV by 8.3%. At steady state, signal changes predominantly demonstrated either NHA or third-harmonic changes. Combined signal analysis improved the PPV by 36%. This method enhanced the detection of thromboembolism and pump thrombosis in the HVAD.Keywords
This publication has 19 references indexed in Scilit:
- Eighth annual INTERMACS report: Special focus on framing the impact of adverse eventsThe Journal of Heart and Lung Transplantation, 2017
- Acoustic Characterization of Axial Flow Left Ventricular Assist Device Operation In Vitro and In VivoASAIO Journal, 2016
- Identification and Management of Pump Thrombus in the HeartWare Left Ventricular Assist Device SystemJACC: Heart Failure, 2015
- Stroke After Left Ventricular Assist Device Implantation: Outcomes in the Continuous-Flow EraThe Annals of Thoracic Surgery, 2015
- Acoustic Spectral Analysis for Determining Pump Thrombosis in Rotary Blood PumpsASAIO Journal, 2014
- Sixth INTERMACS annual report: A 10,000-patient databaseThe Journal of Heart and Lung Transplantation, 2014
- Acoustic Analysis of a Mechanical Circulatory SupportArtificial Organs, 2013
- Asymmetric Pattern of Cerebrovascular Lesions in Patients After Left Ventricular Assist Device ImplantationStroke, 2012
- Continuous Flow Left Ventricular Assist Device Improves Functional Capacity and Quality of Life of Advanced Heart Failure PatientsJACC, 2010
- Neurological Events During Long-Term Mechanical Circulatory Support for Heart FailureCirculation, 2004