CD4-T cell enumeration in human immunodeficiency virus (HIV)-infected patients: A laboratory performance evaluation of Muse Auto CD4/CD4% system by World Health Organization prequalification of in vitro diagnostics
Open Access
- 23 January 2019
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 14 (1) , e0209677
Abstract
CD4 T-cell counts are still widely used to assess treatment eligibility and follow-up of HIV-infected patients. The World Health Organization (WHO) prequalification of in vitro diagnostics requested a manufacturer independent laboratory evaluation of the analytical performance at the Institute of Tropical Medicine (ITM) Antwerp, Belgium, of the Muse Auto CD4/CD4% system (Millipore), a new small capillary-flow cytometer dedicated to count absolute CD4-T cells and percentages in venous blood samples from HIV-infected patients. Two hundred and fifty (250) patients were recruited from the HIV outpatient clinic at ITM. Accuracy and precision of CD4 T cell counting on fresh EDTA anticoagulated venous blood samples were assessed in the laboratory on a Muse Auto CD4/CD4% system. Extensive precision analyses were performed both on fresh blood and on normal and low stabilized whole blood controls. Accuracy ((bias) was assessed by comparing results from Muse CD4/CD4% to the reference (single-platform FACSCalibur). Clinical misclassification was measured at 500, 350, 200 and 100 cells/μL thresholds. Intra-assay precision was < 5%, and inter-assay was < 9%. CD4 T cell counts measured on Muse Auto CD4/CD4% System and on the reference instrument resulted in regression slopes of 0.97 for absolute counts and 1.03 for CD4 T cell percentages and a correlation coefficient of 0.99 for both. The average absolute bias as compared to the reference was negligible (4 cells/μL or 0.5%). The absolute average bias on CD4 T cell percentages was < 1%. Clinical misclassification at different CD4 T cell thresholds was small resulting in sensitivities and specificities equal or >90% at all thresholds except at 100 cells/μL (sensitivity = 87%). All samples could be analyzed as there was no repetitive rejection errors recorded. The Muse Auto CD4/CD4% System performed very well on fresh venous blood samples and met all WHO acceptance criteria for analytical performance of CD4 technologies.Keywords
Funding Information
- World Health Organization
This publication has 14 references indexed in Scilit:
- Thirty‐five years of CD4 T‐cell counting in HIV infection: From flow cytometry in the lab to point‐of‐care testing in the fieldCytometry Part B: Clinical Cytometry, 2016
- Sustainable HIV treatment in Africa through viral-load-informed differentiated careNature, 2015
- The future role of CD4 cell count for monitoring antiretroviral therapyThe Lancet Infectious Diseases, 2015
- WHO Multicenter Evaluation of FACSCount CD4 and Pima CD4 T-Cell Count SystemsJAIDS Journal of Acquired Immune Deficiency Syndromes, 2014
- CD4 cell count and viral load monitoring in patients undergoing antiretroviral therapy in Uganda: cost effectiveness studyBMJ, 2011
- Guidelines for performing single-platform absolute CD4+ T-cell determinations with CD45 gating for persons infected with human immunodeficiency virus. Centers for Disease Control and Prevention.2003
- Performance evaluation of the FACSCount system: A dedicated system for clinical cellular analysisCytometry, 1996
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- A New Biometrical Procedure for Testing the Equality of Measurements from Two Different Analytical Methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part IClinical Chemistry and Laboratory Medicine (CCLM), 1983
- A Coefficient of Agreement for Nominal ScalesEducational and Psychological Measurement, 1960