Relationship between blood pressure obtained from the upper arm with a cuff-type sphygmomanometer and central blood pressure measured with a catheter-tipped micromanometer
- 26 November 2007
- journal article
- research article
- Published by Springer Nature in Heart and Vessels
- Vol. 22 (6) , 410-415
Abstract
Recently, the importance of central blood pressure for cardiovascular risk stratification has been emphasized. Accordingly, the differences in peak systolic and bottom diastolic pressures between the ascending aorta and the brachial artery should be clarified. Study subjects consisted of 82 consecutive patients with suspected coronary artery disease who underwent cardiac catheterization, and in whom ascending aortic pressure waveform was obtained using a catheter-tipped micromanometer, and at the same time systolic and diastolic pressures were measured (single measurement) from the right upper arm with a cuff-type sphygmomanometer based on the oscillometric technique. No significant systematic difference (bias) was found between the peak pressure obtained in the ascending aorta and the systolic pressure from the right upper arm (133.6 ± 25.1 vs 131.8 ± 21.5 mmHg, not significant). Bland–Altman analysis showed only a small bias of +1.8 mmHg, and the limits of agreement were 25.4 mmHg and −21.8 mmHg. In contrast, the bottom pressure in the ascending aorta was significantly lower compared with the diastolic pressure from the upper arm (68.5 ± 10.7 vs 73.0 ± 12.4 mmHg, P < 0.0001). Bland–Altman analysis showed a small but significant bias of −4.5 mmHg, and the limits of agreement were 14.1 mmHg and −23.1 mmHg. The observed biases seemed to remain within practical range. However, random variation in the two measurements was rather large. This is considered to be caused by the random error in the single measurement with the cuff-type sphygmomanometer.Keywords
This publication has 22 references indexed in Scilit:
- Letter by Williams and Tabas Regarding Article “Atherosclerosis 2005: Recent Discoveries and Novel Hypotheses”Circulation, 2006
- Noninvasive Assessment of Local Pulse PressureHypertension, 2005
- Clinical Utility of Aortic Pulses and Pressures Calculated From Applanated Radial-Artery PulsesHypertension, 2003
- Estimation of central aortic pressure by SphygmoCor® requires intra-arterial peripheral pressuresClinical Science, 2003
- Peripheral blood pressure measurement is as good as applanation tonometry at predicting ascending aortic blood pressureJournal Of Hypertension, 2003
- Prospective Evaluation of a Method for Estimating Ascending Aortic Pressure From the Radial Artery Pressure WaveformHypertension, 2001
- National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers.Hypertension, 1993
- Wave reflection in the systemic circulation and its implications in ventricular functionJournal Of Hypertension, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Disparities Between Aortic and Peripheral Pulse Pressures Induced by Upright Exercise and Vasomotor Changes in ManCirculation, 1968