Lateral resolution enhancement of laser scanning microscopy by a higher-order radially polarized mode beam
- 4 August 2011
- journal article
- Published by Optica Publishing Group in Optics Express
- Vol. 19 (17) , 15947-15954
Abstract
We demonstrate that the lateral resolution of confocal laser scanning microscopy is dramatically improved by a higher-order radially polarized (HRP) beam with six concentric rings. This beam was generated simply by inserting liquid crystal devices in front of an objective lens. An HRP beam visualized aggregated 0.17 μm beads individually and is also applicable to biological imaging. This method can extend the capability of conventional laser scanning microscopes without modification of the system, with the exception of the addition of the liquid crystal devices in the optical path.Keywords
This publication has 22 references indexed in Scilit:
- Cylindrical vector beams: from mathematical concepts to applicationsAdvances in Optics and Photonics, 2009
- Sharper focal spot formed by higher-order radially polarized laser beamsJournal of the Optical Society of America A, 2007
- Imaging Intracellular Fluorescent Proteins at Nanometer ResolutionScience, 2006
- Sub-diffraction-limit imaging by stochastic optical reconstruction microscopy (STORM)Nature Methods, 2006
- Sharper Focus for a Radially Polarized Light BeamPhysical Review Letters, 2003
- Ultrasmall focusing spot with a long depth of focus based on polarization and phase modulationOptics Letters, 2003
- Roles of polarization, phase and amplitude in solid immersion lens systemsOptics Communications, 2001
- Focusing light to a tighter spotOptics Communications, 2000
- Surpassing the lateral resolution limit by a factor of two using structured illumination microscopyJournal of Microscopy, 2000
- Breaking the diffraction resolution limit by stimulated emission: stimulated-emission-depletion fluorescence microscopyOptics Letters, 1994