Revision Cochlear Implantation Surgery in Adults
Authors:
- Andrea L. MarloweAMAndrea L. Marlowe
- Jill E. ChinniciJCJill E. Chinnici
- John K. NiparkoJNJohn K. Niparko
- Howard W. FrancisHFHoward W. Francis
- 1 August 2008
- journal article
- Published by Wolters Kluwer Health in Otology & Neurotology
- Vol. 29 (5) , 639-648
Abstract
To assess the efficacy, risks, and indications of revision cochlear implantation (RCI) and to identify the clinical, audiologic, and device-related characteristics that predict outcome. Retrospective case series. Academic tertiary referral center. Adults (> or = 18 yr) who underwent RCI at Johns Hopkins University. Revision cochlear implant surgery. Speech perception by open-set testing and patient report and patient report of symptom resolution. During the 16-year period of this study, 4.8% of all adults implanted at our center have required 1 or more RCI surgeries. A total of 48 RCIs have been performed. The indications for RCI included infection (12%), electrode extrusion (15%), hard failure (23%), suspected device failure (42%), and isolated facial nerve stimulation (8%). Overall, successful resolution of the implant-related or medical condition was achieved with RCI in 83% of cases. Speech perception was lower in only 1 (2.1%) of 48 cases. Satisfactory preoperative speech recognition was preserved or surpassed in 5 of 6 infection cases and 8 cases with progressive symptoms of tinnitus and facial nerve stimulation. All cases of hard failure regained or surpassed previous peak performance. Improved speech recognition was experienced by 75% of cases with suspected device failure. Of cases in which RCI failed to restore previous functional benefit, there was a significant association with advanced age (> 70 yr). Whereas an abnormal integrity test was predictive of favorable outcome after RCI, a negative test was not predictive of outcome. A similar pattern of results was observed with respect to ex vivo device analysis. Revision cochlear implantation can be safely performed to restore lost benefit in appropriately selected cases. When properly performed after medical and audiologic options have been exhausted, RCI rarely compromises previous function and, in most cases, can resolve functional complaints and distracting symptoms. When positive, integrity testing is a useful screen for the presence of a device defect. In cases in which device integrity is uncertain, clinical judgment guided by longitudinal assessment can help determine whether RCI is likely to be beneficial.Keywords
This publication has 26 references indexed in Scilit:
- Cochlear Reimplantation: Causes of Failure, Outcomes, and Audiologic PerformanceThe Laryngoscope, 2007
- Quality Control After Insertion of the Nucleus Contour and Contour Advance Electrode in AdultsEar & Hearing, 2007
- A Multicenter Study of Device Failure in European Cochlear Implant CentersEar & Hearing, 2007
- Medical and surgical complications of the second 100 adult cochlear implant patients in BirminghamThe Journal of Laryngology & Otology, 2005
- Cochlear Implant Soft Failures Consensus Development Conference StatementOtology & Neurotology, 2005
- Reliability of cochlear implantsOtolaryngology–Head and Neck Surgery, 2005
- Is Word Recognition Correlated With the Number of Surviving Spiral Ganglion Cells and Electrode Insertion Depth in Human Subjects With Cochlear Implants?The Laryngoscope, 2005
- Cochlear Reimplantation: Surgical Techniques and Functional ResultsThe Laryngoscope, 2001
- Cochlear reimplantationThe Laryngoscope, 1999
- Lymphangioma presenting as a delayed posttraumatic expanding neck massAmerican Journal of Otolaryngology, 1995