The Technicon AutoAnalyzer method, the Beckman Creatinine Analyzer II, DuPont aca, Electro-Nucleonics GEMSAEC and instrumentation Laboratory Model 919 kinetic methods for determination of [human] serum creatinine were evaluated. These methods, which involve the Jaffe reaction, all showed a statistically significant proportional error relative to an ion-exchange pretreatment method for creatinine in 92 serum samples; the GEMSAEC method also showed a bias. The positive interference of several acid metabolites was quantified in each method. Bilirubin, although associated with a negative creatinine interference, is apparently not the sole interferent. The magnitude of this interference was determined and several methods for minimizing it were studied.