Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study
Open Access
- 25 September 2020
- journal article
- research article
- Published by Springer Nature in Radiation Oncology
- Vol. 15 (1) , 1-13
Abstract
Background: Rotational beam delivery enables concurrent acquisition of cone-beam CT (CBCT), thereby facilitating further geometric verification of patient setup during radiation treatment. However, it is challenging to acquire CBCT during stereotactic body radiation therapy (SBRT) using flattening-filter free X-ray beams, in which a high radiation dose is delivered. This study presents quantitative evaluation results of the image quality in four-dimensional (4D) in-treatment CBCT acquired during SBRT delivery. Methods: The impact of megavoltage (MV) scatter and acquisition parameters on the image quality was evaluated using Catphan 503 and XSight lung tracking phantoms. The in-treatment CBCT images of the phantoms were acquired while delivering 16 SBRT plans. The uniformity, contrast, and contrast-to-noise ratio (CNR) of the in-treatment CBCT images were calculated and compared to those of CBCT images acquired without SBRT delivery. Furthermore, the localizing accuracy of the moving target in the XSight lung phantom was evaluated for 10 respiratory phases. Results: The CNR of the 3D-reconstucted Catphan CBCT images was reduced from 6.3 to 2.6 due to the effect of MV treatment scatter. Both for the Catphan and XSight phantoms, the CBCT image quality was affected by the tube current and monitor units (MUs) of the treatment plan. The lung target in the XSight tracking phantom was most visible for extreme phases; the mean CNRs of the lung target in the in-treatment CBCT images (with 40 mA tube current) across the SBRT plans were 3.2 for the end-of-exhalation phase and 3.0 for the end-of-inhalation phase. The lung target was localized with sub-millimeter accuracy for the extreme respiratory phases. Conclusions: Full-arc acquisition with an increased tube current (e.g. 40 mA) is recommended to compensate for degradation in the CBCT image quality due to unflattened MV beam scatter. Acquiring in-treatment CBCT with a high-MU treatment beam is also suggested to improve the resulting CBCT image quality.Keywords
Funding Information
- Elekta
This publication has 28 references indexed in Scilit:
- Verification of Planning Target Volume Settings in Volumetric Modulated Arc Therapy for Stereotactic Body Radiation Therapy by Using In-Treatment 4-Dimensional Cone Beam Computed TomographyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2013
- Target‐specific optimization of four‐dimensional cone beam computed tomographyMedical Physics, 2012
- A megavoltage scatter correction technique for cone-beam CT images acquired during VMAT deliveryPhysics in Medicine & Biology, 2012
- Mitigating Errors in External Respiratory Surrogate-Based Models of Tumor PositionInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- A novel method for megavoltage scatter correction in cone-beam CT acquired concurrent with rotational irradiationRadiotherapy and Oncology, 2011
- First clinical cone-beam CT imaging during volumetric modulated arc therapyRadiotherapy and Oncology, 2009
- Verification of In-Treatment Tumor Position Using Kilovoltage Cone-Beam Computed Tomography: A Preliminary StudyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Improving IMRT dose accuracy via deliverable Monte Carlo optimization for the treatment of head and neck cancer patientsMedical Physics, 2006
- Correlation of lung tumor motion with external surrogate indicators of respirationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- Detection of lung tumor movement in real-time tumor-tracking radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2001