Stent imaging on a clinical dual-source photon-counting detector CT system—impact of luminal attenuation and sharp kernels on lumen visibility
- 3 December 2022
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 33 (4) , 2469-2477
Abstract
Objectives To assess the impact of scan modes and reconstruction kernels using a novel dual-source photon-counting detector CT (PCD-CT) on lumen visibility and sharpness of different stent sizes. Methods A phantom containing six balloon-expandable stents (2.5 to 9 mm diameter) in silicone tubing was scanned on a PCD-CT with standard (0.6 mm and 0.4 mm thicknesses) and ultra-high-resolution (0.2 mm thickness) modes. With the use of increasing contrast medium concentrations, densities of 0, 200, 400, and 600 HU were achieved. Standard-resolution scans were reconstructed using increasing sharpness kernels, using both polyenergetic quantitative soft tissue “conventional” ((Qr40c(0.6 mm), Qr40c(0.4 mm), Qr72c(0.2 mm)) and vascular (Bv) virtual monoenergetic reconstructions (Bv44m(0.4 mm), Bv60m(0.4 mm)) at 70 keV. In-stent lumen visibility, sharpness (max. ΔHU of the stent measured in profile plots), and in-stent noise (standard deviation of HU) were measured. Results In-stent lumen visibility was highest for Qr72c(0.2 mm) (86.5 ± 2.8% to 88.3 ± 2.6%) and in Bv60m(0.4 mm) reconstructions (77.3 ± 2.9 to 82.7 ± 2.5%). Lumen visibility was lowest in the smallest stent (2.5 mm) ranging from 54.1% in Qr40c(0.6 mm) to 74.1% in Qr72c(0.2 mm) and highest in the largest stent (9 mm) ranging from 93.8% in Qr40c(0.6 mm) to 99.1% in the Qr72c(0.2 mm) series. Lumen visibility decreased by 2.1% for every 200-HU increase in lumen attenuation. Max. ΔHU between stents and stent lumen was highest in Qr72c(0.2 mm) (ΔHU 892 ± 504 to 1526 ± 517) and Bv60m(0.4 mm) series (ΔHU 480 ± 357 to 1030 ± 344). Improvement of lumen visibility and sharpness in UHR and Bv60m(0.4 mm) series was strongest in smaller stent sizes. Conclusion UHR acquisition mode and sharp reconstruction kernels on a novel PCD-CT system significantly improve in-stent lumen visibility and sharpness—especially for smaller stent sizes. Key Points • In-stent lumen visibility and sharpness of stents significantly improve using sharp reconstruction kernels (Bv60) and ultra-high-resolution mode in photon-counting detector computed tomography. • The observed improvement of stent-lumen visibility was highest in smaller stent sizes.Keywords
Funding Information
- Siemens Healthineers
This publication has 32 references indexed in Scilit:
- Photon-Counting CTInvestigative Radiology, 2018
- Optimized energy of spectral coronary CT angiography for coronary plaque detection and quantificationJournal of Cardiovascular Computed Tomography, 2018
- The Updated NICE Guidelines: Cardiac CT as the First-Line Test for Coronary Artery DiseaseCurrent Cardiovascular Imaging Reports, 2017
- Comprehensive Comparison of Virtual Monoenergetic and Linearly Blended Reconstruction Techniques in Third-Generation Dual-Source Dual-Energy Computed Tomography Angiography of the Thorax and AbdomenInvestigative Radiology, 2016
- Human Imaging With Photon Counting–Based Computed Tomography at Clinical Dose LevelsInvestigative Radiology, 2016
- Next generation coronary CT angiography: in vitro evaluation of 27 coronary stentsEuropean Radiology, 2014
- Diagnostic accuracy of in-stent coronary restenosis detection with multislice spiral computed tomography: a meta-analysisEuropean Radiology, 2007
- Assessment of coronary artery stent restenosis by 64-slice multi-detector computed tomographyEuropean Heart Journal, 2006
- 64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stentsEuropean Radiology, 2005
- Imaging of coronary artery stents using multislice computed tomography: in vitro evaluationEuropean Radiology, 2003